3. Connecting Pattern
3.1 Testing Ports
The tester has four jacks which connect with A+/V+/V-/A-. A+ and A- ports are connected with
the positive and negative poles of the test objects, working as the current channel in
discharging. V+ and V- ports are connected with the positive and negative poles of the test
objects, working as the voltage channel in voltage tests. This way of using four individual
cables will help eliminate the effect of wires on the test result and improve the test precision.
3.2 PC Connection
The tester can be connected with a computer through a MicroUSB cable.
3.3 Input Power:
The tester will work on the power supplied through MicroUSB port.
4. Display and Setting
4.1 Buttons
SET —— setting knob (rotate to left/right and press)
4.2 Operation Display
The tester can show voltage, current and capacity during working. You can switch to different
data by rotating “SET” knob.
Voltage display: Automatically switch between 0.00V to 00.0V
Current display: 0.00A
Capacity display: Auto switch between 0.000 and 00.00Ah.
Energy display: Auto switch between 0.000 and 00.00Wh.
Press on “SET” knob to start test under testing mode and press again to stop test..
You can open setting mode by pressing on the “SET” knob for 2 seconds in the stop state (This
function is invalid when the tester being connected with the PC software).
4.3 Setting Display
In setting mode, the value to be set will shine. Press on “SET” knob to switch to the value you
want to change and rotate “SET” knob to set the value.
First set the discharging current: you can set the value of every digital.
Second set the cut-off voltage: testing will automatically stop when the testing voltage reaches
below this value.