EBC-A01 User Manual
Copyright (C) 2006-2017 ZKETECH
The second line shows constant voltage, cut-off current (charging stops automatically when
testing current reaches below this parameter, it could be set above 0.1A) and auto mode
settings:
²
NOR
–
normal mode, testing stops when charging completes.
²
AUTO
–
automatically complete the
“
charging-discharging-charging
”
process.
When the cursor is on AUTO, a long press on
“
ON
”
and you will open the second step of
automatic discharging and it allows you to set discharging parameters.
The first parameter: discharging current; the second parameter: cut-off voltage; the third
parameter: waiting time between step switches (a 5-10 minute waiting time is suggested
between charging and discharging to cool the battery).
After completes settings, press
“
SET
”
button for a long time to save data and return to the
upper-level interface.
Press
“
SET
”
button for a long time in the main setting interface to save data and reture to the
testing interface. In the testing interface press
“
ON
”
button to start testing.
4.4 Setting examples
1) Setting steps
Ø
Turn on the battery tester.
Ø
Connect battery and you can read the battery voltage on the testing interface on the
screen.
Ø
Press
“
SET
”
button for more than two seconds to open the setting interface.
Ø
Set testing parameters.
Ø
Press
“
SET
”
button for more than two seconds to return to testing interface.
Ø
Press
“
ON
”
button to start testing.
Ø
When testing completes you can read results in the testing interface.
Ø
Remove the battery and turn off the tester.
2) Setting discharging at a constant current (discharging a button cell battery to 2V at 0.005A)
3) Charging at a constant current (charging 3.7V LiPo to 4.2V at 0.1A, cut-off current sets
0.05A)