Oxygen Monitor SGM7 5 Technical data
HB_SGM7_eng.docx
13
Data bits ........................................ 8
Parity ............................................. None
Handshake .................................... None
Pin assignment connector SUB-D 9-pol. F:
Chart 1:
Serial interface protocol (CR = Carriage Return)
Command
Display
Description
M2CR
M2x.xxExxCR
Oxygen concentration in ppm
A1CR
A1xxxCR
Cell voltage in mV
A2CR
A2xxxCR
Measuring cell temperature in °C
Chart 2:
Error indication
Error code
Description
Error0
Transmission error
Error1
Warm up
Error2
Cell temperature too low
Error3
Thermocouple defective
Error6
System error
Analog output
Current signal ......................... 0/4 - 20 mA, potential-free, range adjustable,
apparent Ohmic resistance max. 500 Ohm
Alternative voltage signal ........ 0/2 - 10V (on request), potential-free,
range adjustable
Chart 3:
Pin assignment for 4-pol. socket
Pin
Description
1
I
out
– (optional U
out
– )
2
I
out
+ (optional U
out
+ )
3
Alarm relay (potential-free contact)
4
Alarm relay (potential-free contact)
Chart 4:
Limit value relay
Charge
Resistive charge (cos
ϕ
= 1)
Max. switching voltage
125 V AC, 60 V DC
Max. switching current
1A
Max. switching power
62.5 VA, 30 W
Min. permissible charge
1 mA at 5 V DC
MD
TXD
RXD