www.zinexts.com                                                                                                              MagPurix 24 EVO User Manual

 

4

  / 

35

 

 Introduction 

1.1

 

About the MagPurix 24 EVO System 

Thank you for purchasing the MagPurix 24 EVO System for your laboratory. This guide contains 
important  information  regarding  the  safe  use of  the

  MagPurix  24  EVO

  System. Please  read  this 

manual carefully, before starting to operate the instrument for the first time, especially the Safety 
Information.

 

If there is any question about how to install or operate the system, please contact our certified 
distributors/agents or email our technical support center ([email protected]).

 

 

1.2

 

Technology Introduction 

1.2.1

 

Principle 

The MagPurix 24 EVO System is a fully automated clinical sample preparation system that enables 
performing purification of nucleic acids from a wide range of sample types for molecular diagnostic 
in the area of infectious diseases, human health, personalized medicine, and life sciences research.

 

The MagPurix 24 EVO System is designed for the automated extraction and isolation of nucleic acids, 
utilizing advanced magnetic bead separation technology.

 

1.2.2

 

Applications 

The  MagPurix 24 EVO  System is intended for  in  vitro  diagnostic  (IVD)  use, for  performing DNA 
and/or  RNA  extraction,  from  a  variety  of  samples  which  are  essential  for  clinical  diagnostics, 
personalized medicine, forensics, etc. 

 

 

1.3

 

Safety Information 

This chapter presents safety information that you need to be familiar with and strictly follow before 
you  operate,  maintain  or  service  this  instrument.  If  you  fail  to  follow  the  safety  information,  or 
neglect the warning in this user manual, it might cause damage to the instrument. Please note that 
operating the instrument outside its intended use might lead to personnel injury and material loss.

 

For your safety and those of others, follow carefully the guidelines provided in the following pages 
concerning the use of the MagPurix 24 EVO System.

 

 

Zinexts Life Science Corp. will not be held liable for any errors, damage, or other unexpected events 
resulting from misconduct on the following safety guidelines.

 

Summary of Contents for MagPurix 24 EVO

Page 1: ...rp User Manual Automated Nucleic Acid Purification System 24 EVO Catalog No ZP01019 Manual No QW IH01 20 Version 3 0 Obelis s a Bd G n ral Wahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 73...

Page 2: ...Safety Symbols Labels 5 1 3 2 About the Instrument 5 1 3 3 About Reagent Kits 6 1 3 4 About Samples 7 1 3 5 About Infectious Waste 7 2 Intended Use General Description 8 2 1 Intended Use 8 2 2 Externa...

Page 3: ...Operation 19 5 1 Preparation 19 5 2 Start up the System 20 5 2 1 Basic Settings 20 5 2 2 User Management 20 5 2 3 Door Lock and Door Sensor Switch 20 5 3 Perform Extraction 21 5 4 System Fixing before...

Page 4: ...www zinexts com MagPurix 24 EVO User Manual 3 35 9 2 Technical Support 34 10 Warranty Information 34 Revision History 35...

Page 5: ...gPurix 24 EVO System is designed for the automated extraction and isolation of nucleic acids utilizing advanced magnetic bead separation technology 1 2 2 Applications The MagPurix 24 EVO System is int...

Page 6: ...indicates that certain precautions must be taken when working with potentially infectious material Hot Surface This symbol labels potentially hot surfaces on the instrument Electricity This symbol is...

Page 7: ...bjects on the MagPurix 24 EVO System Also refrain from bumping or knocking it as doing so may cause a failure or malfunction of the instrument If any liquid materials are left or have leaked inside th...

Page 8: ...3 5 About Infectious Waste Caution When handling or disposing of infectious materials follow the laboratory guidelines or the law regarding infectious waste to perform proper incineration fusion steri...

Page 9: ...ple types The MagPurix 24 EVO System is intended to be used only in combination with extraction kits indicated for use with MagPurix System for the applications described in the respective kit handboo...

Page 10: ...www zinexts com MagPurix 24 EVO User Manual 9 35 2 3 Internal Features...

Page 11: ...arts MagPurix 24 EVO System x 1 Power Cord x 1 Sample Rack for 2 ml Sample Tube x 2 Barcode Reader x 1 User Manual x 1 SD Card 32GB x 1 64GB x 1 Fuse x 3 1A 2A 4A Please make sure all components are f...

Page 12: ...up and down suddenly Warming a cold room suddenly or moving the MagPurix 24 EVO System from a room with low temperature to a warm room may cause condensation inside the device resulting in abnormal op...

Page 13: ...might lead to instrument damages For correct and safe use of the MagPurix 24 EVO System install the device in a location close to a power socket and with enough space for installation procedures and...

Page 14: ...required for extraction are provided with the respective extraction reagent kits Contents of the kits will vary please refer to the resp extraction kit handbook for details 3 3 4 User Training User t...

Page 15: ...wering up the MagPurix 24 EVO System the MagPurix 24 EVO System will start the initialization process Click LOGIN and User ID and Password will appear for input Set up the local time System will start...

Page 16: ...nual 15 35 Camera time is suggested to be set for the first time use Follow the steps below to set the time Step 1 Remove the Camera as below Step 2 Turn on the instrument and click on Settings System...

Page 17: ...35 Step 4 Find the buttons from the left side of the camera Step 5 Camera will start recording automatically after sensor is turned off Press OK button to pause Step 6 Long press M button to access to...

Page 18: ...ser Manual 17 35 Step 8 Select Date Time and press OK Step 9 Use the button to set the time and press OK Step 10 Press M button to return to the main menu Step 11 On the panel click CAMERA to turn on...

Page 19: ...access to users to certain options Once logged in you will be able to set up at least 10 other users with their own passwords 4 2 Main Screen The Main Screen of the MagPurix 24 EVO system on the righ...

Page 20: ...he processed protocols and or some samples used with this instrument may contain infectious agents Caution Before start operating the instrument put on appropriate gloves a mask and safety goggles if...

Page 21: ...are version installed is displayed at the bottom of the screen by clicking the Status Bar 5 2 2 User Management After logging in as administrator the Main Screen opens Click the Settings button and on...

Page 22: ...MagPurix 24 EVO Software the Main Screen opens Click the Perform Run button and on the Perform Run Screen click the Extraction button Place reaction cartridges and all plastic consumables onto the wor...

Page 23: ...er to separate reagent cartridge strips reagent chambers and tip holders if you do not run full batches of samples The position of Piercing Pin on the right and Filter Tip on the left are as follows T...

Page 24: ...www zinexts com MagPurix 24 EVO User Manual 23 35 Note Zinexts does not guarantee with expired kit performance...

Page 25: ...al To amend ID information DEL the wrong ID information and then scan the right ID Please take the Sample Tube Rack with the eluates and the samples out of the extraction module and discard all used e...

Page 26: ...and 75 ethanol Clean the heating module with a soft paper tissue or soft cloth and use a cotton swab to remove any potential buffer residue from the hold of the heating module Run disinfection by tur...

Page 27: ...tatus of the piston units Cleaning of piston UV light change annually Service Engineer 6 1 Cleaning of Rubbish Tray and M Motor Elevate M motor before the start of cleaning Clean the rubbish tray and...

Page 28: ...24 EVO User Manual 27 35 6 2 Cleaning of Piston Enter Settings menu to select System and then click Piston Clean piston with 75 ethanol using soft cloth Change new piston if there is significant impur...

Page 29: ...urns on when the power is on 1 The packing materials are not removed 2 Technical problem 1 Turn off the power and remove the packing materials 2 Contact your local representative or agent Protocol sto...

Page 30: ...eparation of beads Too much sample material Reduce amount of sample material to the values recommended in the resp extraction kit handbook Carryover of magnetic beads into eluate Too much sample mater...

Page 31: ...leakage from the tips 1 Perform preventive maintenance annually to ensure proper motor movements 2 Re load the consumables according to the instructions in this manual Low DNA yield Incomplete lysis D...

Page 32: ...contain any particulate material that can clog the tip If needed centrifuge the sample prior to the extraction process No DNA recovered Magnetic beads stored or handled improperly Store cartridge cont...

Page 33: ...event bubble formation during mixing make sure the sample volume is at least at the recommended volume listed in the resp extraction kit handbook Purified DNA repeatedly frozen and thawed Aliquot puri...

Page 34: ...0 C assuming the room temperature at 25 C Protocol Input Touch Panel UV Light 30 minutes at 250 nm Built in Features 7 Touch Panel Instrument Dimensions 91 5 cm W x 66 5 cm D x 60 4 cm H Weight 100 kg...

Page 35: ...at each product described herein will be free from defects in materials and workmanship for a period of twelve 12 months from the date of delivery Zinexts Life Science Corp agrees as its sole responsi...

Page 36: ...ding on or obligate Zinexts Life Science Corp in any way In no event will Zinexts Life Science Corp be liable for any special incidental or consequential damages resulting from use or malfunction of t...