3.0 Cleaning &Maintaining the MagPurix 12A
system
Two types of maintenance are performed on the MagPurix 12A system
Instrument as listed in the table. For details on each type of maintenance,
see below:
Schedule
Maintenance Type
Performed by
After each use
Routine
Cleaning of Sample Tray, Piston
M module by 75% EtOH.
Cleaning of heating module
and using cotton swab to check
buffer residue in hold of heating
module.
User
Bi-weekly
Cleaning of instrument body
(outside)
Cleaning of instrument body
(inside)
User
Two times a year or
Annually
Preventive
Grease ball screws if required
Check motor movement
Check the air-tight status with
piston units (using empty plastic
wares and run )
Service Engineer
Clean the Sample Tray with mild detergent and rinse with
deionized water. Allow the parts to dry before use. Clean and
disinfect the platform surface by wiping with deionized water
followed by 75% ethanol.
Clean the instrument body by removing dust gently with a dry,
soft cloth. If the outside of the MagPurix 12A system is heavily
soiled, or if any samples that may cause infection are adhered to
the outside of the MagPurix 12A system, wipe with a soft paper
tissue, soaked with 75% ethanol.