SECTION
10
Instruments
NexGen
Posterior Referencing
Instruments – Tibial Technique
Position Tibial Pin Guide
•
Remove the meniscus.
•
Do not remove osteophytes from the tibia. Clean the bone
around the ACL and anterior tibia where the guide fits.
•
Remove any soft tissues that may prevent proper placement
of the Pin Guide.
•
Position the Pin Guide on the tibia
(Fig. 23)
.
•
The rotation cylinders do not make contact with bone surface.
Both cylinders will be above the tibia surface. All other
contact with the Pin Guide should ensure a proper fit.
•
The mechanical axis and proximal resection reference lines
on the Pin Guide can be used to assess alignment.
Drill and Pin Anterior Pin Holes
•
Drill and pin the medial and lateral anterior holes of the
Pin Guide
(Fig. 24)
.
Fig. 23
Fig. 24
Instruments
PSI Tibia
Pin Guide
00-5970-015/
018-00
Universal Pin
and Drill Kit
2001-00-000
NexGen
®
Posterior Referencing Instruments – Tibial Technique
2
B