Zimmer 8831-001 Instruction And Service Manual Download Page 2

WARNING AND CAUTION LABELS

See Fig. 3.

SYMBOL DEFINITIONS

See Fig. 4.

Zimmer

TISSUE BANK Dermatome 

Components (REF 8831)

1. Power Supply (REF 8831-06)

2. Handpiece with cord (REF 8831-01)

3. Screwdriver (REF 8803)

4. Autoclave case (REF 8831-03)

5. Width plate, 3 in. (7.6 cm) (REF 8831-04)

6. Width plate, 4 in. (10.2 cm) (REF 8831-05)

7. Fuse drawer, 5 mm x 20 mm

8. Fuses, 1.25 A (Qty 2) (Installed)

9. Fuses, 1.25 A (Qty 2) (5 mm x 20 mm included)

10. Power Cord

11. Instruction and Service Manual

ZIMMER

TISSUE BANK DERMATOME 

ILLUSTRATIONS

See Fig. 5–7.

BEFORE OPERATION

Connect the 

Zimmer

Tissue Bank Dermatome to its power

supply by inserting the connector plug into the connector
receptacle. (See Fig. 8.) Disconnect the power supply from
the power source before inserting. Remove any kinks or
twists from the cord. Align the connector plug to the
receptacle by rotating the plug while gently pushing inward.
When aligned, push in until the plug clicks into the
receptacle.

To remove the connector plug, pull outward on the outer
collar portion of the plug to separate the plug from the
receptacle.

To replace input fuses, disconnect the power supply from
the power source. Remove the fuse drawer by lifting up on
the tab using a tool and pulling the drawer outward. (See
Fig. 9.) Replace the fuse drawer with the correct fuses
installed. The correct fuse type is listed in the SERVICE
INFORMATION section. The unit is shipped with a 0.25 in.
(6.35 mm) x 1.25 in. (31.75 mm) fuse drawer installed. A
separate fuse drawer is supplied for 0.197 in. (5 mm) x
0.787 in. (20 mm) fuses.

Before connecting the power supply to a power source, be 
sure that the switch on the front of the power supply is in 
the 

OFF (O)

position, and that the safety lock is in the

SAFE

position. Connect to a power source via the "IEC"

connector in the rear of the power supply

Turn power on to the power supply by moving the rocker
switch on the front of the power supply to the 

ON (I)

position. The rocker switch should illuminate. Move the
safety lock to the 

ON

position and briefly depress the on/off

lever to ensure that power is being applied to the
handpiece, and that the carriage is oscillating. Return the
safety lock to the 

SAFE

position.

OPERATING PRECAUTIONS

The on/off lever must be in the SAFE position before
changing blades when the instrument is not in use, or
when connecting power to the instrument.

Accidental

activation of the instrument during these procedures may
injure the tissue bank personnel.

To avoid any injury, use extreme caution when handling the
blade.

Use extreme caution when handling the dermatome with 
the blade installed. Before using, always inspect the cord 
for signs of wear or damage. A worn or damaged cord 
must be returned for service immediately for continued 
safety.

Use only 

Zimmer

Tissue Bank Dermatome Blades 

(REF 8831-000-10). 

Zimmer

Tissue Bank Dermatome

Blades have been specifically designed for use with the

Zimmer

Tissue Bank Dermatome. Other blades will not

fit properly in the dermatome and may cause
deeper grafts than selected.

OPERATING INSTRUCTIONS

1. Place the power supply in the 

OFF (O)

position. Place 

the dermatome in the 

SAFE

position by sliding the safety

lock on the on/off lever toward the blade end of the
instrument to the 

SAFE

position. Only the word 

SAFE

should be visible.

2. Choose the proper width plate to satisfy cutting
requirements. Two width plates are supplied: 3 and 4 in. 
(7.6, and 10.2 cm).

3. Install the blade. (Refer to Blade Placement diagram,
(Fig. 7.) Use caution when inserting the blade to avoid
nicking it, which may result in an uneven cut. Use a new
sterile blade for each procedure. 

a.  Place the on/off lever in the 

SAFE

position.

b.  Using the Zimmer screwdriver (REF 8803), loosen the 

width plate screws approximately two turns. DO NOT
remove screws from handpiece.

c.  Remove width plate and used blade.

d.  Place a new blade onto the carriage.

e.  Align the two holes in the blade with the two pins on

the carriage.

f.   Place width plate over blade and tighten screws. 

4. Control lever adjustment determines graft thickness. (See
Fig. 10.) Factory markings indicate 0.004 in. (0.102 mm) 
divisions. Set control knob pointer to desired thickness.

A semi-annual calibration check by an authorized
Zimmer Orthopaedic Surgical Products service center 
is recommended to verify continued accuracy.

5. To activate the dermatome, place the power supply in the

ON (I)

position. Lift the on/off lever and slide the safety lock

back to the 

ON

position. Depress the on/off lever to start 

the motor. 

DO NOT run the dermatome without cutting 

for an extended time

, as this needlessly wears the blade.

- 2 -

Summary of Contents for 8831-001

Page 1: ...receiving the Zimmer Tissue Bank Dermatome examine the unit for external signs of damage Retain all packing material until the contents have been verified and an initial operational check has been pe...

Page 2: ...CAUTIONS The on off lever must be in the SAFE position before changing blades when the instrument is not in use or when connecting power to the instrument Accidental activation of the instrument durin...

Page 3: ...ply Safely dispose of all used blades Thoroughly scrub the dermatome and width plate s with a soft brush and mild detergent All traces of blood coagulated material stains etc should be removed Rinse a...

Page 4: ...amage caused by misuse abuse accident neglect or any uses not prescribed in this manual If the unit becomes defective because of mis use or abnormal conditions of operation repairs will be billed at o...

Page 5: ...5 1 2 A Autoclave Case B Power Supply C Screwdriver D Width Plates E Handpiece A E D C B...

Page 6: ...cle D Power Entry Module E Back SYMBOL DEFINITIONS A Type of Equipment B Direct Current C Year of Manufacture D OFF Power Disconnection from the Mains E On Power Connection to the Mains F Alternating...

Page 7: ...7 Handpiece A On Off Safety Lock B On Off Lever C Handpiece D Connector Plug E Thickness Control Lever Blade Replacement A Screwdriver B Width Plate C Blade D Carriage 6 7 10 9 8 A A B C D B C D E...

Page 8: ...8 11 15 12 METHOD I A Handpiece B Pocket METHOD II A Handpiece B Forceps 13 14 A B A B...

Page 9: ...DESCRIPTION Zimmer Tissue Bank Dermatome Complete includes 8831 01 8821 06 8831 03 REF 8831 06 DESCRIPTION Power Supply REF 8831 03 DESCRIPTION Autoclave Case REF 8831 01 DESCRIPTION Handpiece includ...

Page 10: ...831 000 10 DESCRIPTION Zimmer Tissue Bank Dermatome Blades Sterile 10 per box REF 8831 04 8831 05 DESCRIPTION Width Plates REF 8803 001 10 DESCRIPTION Width Plate Screws 10 per pack REF 8803 DESCRIPTI...

Page 11: ...11...

Page 12: ...Zimmer Orthopaedic Surgical Products 200 West Ohio Avenue P O Box 10 Dover Ohio U S A 44622 Revised 05 05 2005 Zimmer Inc Printed in U S A 18 10620 Printed on Recycled Paper...

Reviews: