Place the throttle in the SAFE position. To place the dermatome in the SAFE position, slide the 
safety lock on the throttle toward the blade end of the instrument to the SAFE position. Only the 
word SAFE should be visible.  

 

• 

Using a screwdriver, loosen width plate screws approximately two turns. Do not remove screws 
from handpiece. 

 

• 

Place a new blade in slot on the handpiece. If replacing a blade, remove the used blade before 
inserting a new one. Refer to BLADE REMOVAL section. 
 

• 

Mate the drive pin with the hole in the blade. Note: “INSERT WITH THIS SIDE UP” message.  

 

• 

Lubrication of the blade is not necessary because the backing of the blade is a self-lubricating 
plastic.  

 

• 

Choose proper width plate to satisfy cutting requirements. Place width plate over blade and 
tighten screws. Do not overtighten. Ensure the printing on the width plate is facing out. (See Fig. 
5.) 

CUTTING THE GRAFT 

 

• Skin should be prepared in routine manner. 

 

• It is not necessary to lubricate the skin; however, lubricating the donor site with sterile mineral oil may 

ease travel of the Zimmer Electric Dermatome.  

• Set control lever adjustment knob pointer to desired graft thickness. Factory calibrations indicate 0.002 
in. (0.050 mm). Do not insert any instrument between the blade and the control lever as this may damage 
or knick the blade causing a poor cut. Further, it may compromise the calibration of the instrument. (See 
Fig. 8.)  

• Hold the handpiece on the donor site at a 30° – 45° angle. (See Fig. 10, 11.) 

 

• To activate the dermatome, place the power supply in the ON (I) position. Lift the throttle lever and slide 

the safety lock back from the SAFE position toward the hose coupling. The word ON should be visible. 
(See Fig. 9.) For optimum results, it is recommended that the dermatome operate at full speed. To ensure 
that full speed is achieved, completely depress the throttle control with the safety lock in the ON position.  

• Depress the throttle to start the cut. Guide the unit forward using a slight downward pressure to ensure 
that the cutting edge remains continuously firmly in contact with the donor site. 

 

• Two methods of graft removal from the instrument may be used: 

• 

Method I  

Summary of Contents for Electric Dermatome

Page 1: ...POWER SUPPLY I Physical Weight 2 6 lb 1 18 kg Length 9 12 in 23 2 cm Height 5 44 in 13 8 cm Width 6 69 in 17 0 cm II Electrical Power Requirements 100 240 V Auto Switching 150 VA 50 60 Hz Single Phase Power Output 14 5 V 4 3 A Maximum Protection Class Class 1 Degree of Protection Against Electrical Shock Type BF II HANDPIECE I Physical Weight 2 1 lb 95 kg Width 8 5 in 21 6 cm Length 5 2 in 13 2 cm...

Page 2: ...FE position the safety lock on the throttle should be toward the blade end of the dermatome and only the word SAFE should be visible Handle the Zimmer Electric Dermatome carefully Should it be inadvertently dropped or damaged it should be serviced Do not use Use only Zimmer Dermatome Blades REF 00 8800 000 10 The Zimmer Dermatome Blade has been specifically designed and engineered for use with the...

Page 3: ...he connector plug to the receptacle by rotating the plug while gently pushing inward When aligned push in until the plug clicks into the receptacle Note Older Zimmer Dermatomes and Power Supplies were equipped with a metal connector plug and receptacle The metal connectors have been upgraded to plastic connectors for ease of use The older metal connectors are not compatible with the plastic connec...

Page 4: ...wever lubricating the donor site with sterile mineral oil may ease travel of the Zimmer Electric Dermatome Set control lever adjustment knob pointer to desired graft thickness Factory calibrations indicate 0 002 in 0 050 mm Do not insert any instrument between the blade and the control lever as this may damage or knick the blade causing a poor cut Further it may compromise the calibration of the i...

Page 5: ...sue with a disposable non shedding wipe and cover with a damp cloth Body fluids and tissue should not be allowed to dry on instruments prior to cleaning Return and secure instrument and width plates to instrument case for transport and cleaning BLADE REMOVAL Loosen the width plate screws Do not remove the screws To remove the width plate hold both sides and lift Do not lift the width plate from th...

Page 6: ...cautions for handling contaminated biohazardous materials should be observed Instruments should be cleaned within 30 minutes of use to minimize the potential for drying prior to cleaning Prepare cleaning agents at the use dilution and temperature recommended by the manufacturer Thoroughly scrub the instrument with a soft bristled brush and a pH neutral detergent If desired a neutral pH enzyme solu...

Page 7: ...outgassing times are difficult to determine for lubricated powered instruments Never sterilize the power supply Disconnect the dermatome from the power supply before sterilization Place cleaned instruments in an instrument tray or fully perforated autoclave case The Zimmer Dermatome Autoclave Case REF 00 8801 003 00 is recommended See Fig 18 Do not kink or crimp the dermatomeʼs cord when closing t...

Page 8: ... table 4 Disinfection steam sterilization parameters recommended by the World Health Organization WHO for reprocessing instruments where there is concern regarding TSE CJD contamination 5 For Universal Instrument Cases without defined load configurations 6 Sea level 7 AAMI AORN steam sterilization cycles with longer times than those listed are also acceptable 8 Medical grade steam sterilization co...

Page 9: ...00 8802 003 00 00 8802 004 00 00 8803 000 00 REF 00 8821 001 00 Description Power Supply includes Fuse drawer 5 mm x 20 mm Fuses 1 25 A QTY 2 Fuses 800 mA QTY 2 Power Cord REF 00 8821 006 00 Description Autoclave Case REF 00 8801 003 00 Description 1 in 2 5 cm Width Plate REF 00 8802 001 00 Description 1 5 in 3 8cm Width Plate REF 00 8802 015 00 Description 2 in 5 1 cm Width Plate REF 00 8802 002 ...

Page 10: ...tions Have unit serviced Dermatome not operating Power light on Check blade fit Check handpiece connection to power supply Have unit serviced Dermatome operating too slowly Check dermatome for damage Check blade for damage Have unit serviced Erratic speed changes Check handpiece connection to power supply Have unit serviced SERVICE INFORMATION Annual calibration checks are strongly recommended to ...

Page 11: ...A Autoclave Case B Power Supply C Screwdriver D Width Plates E Handpiece ...

Page 12: ...A Front B Mains Switch C Connector Receptacle D Power Entry Module E Back A On Off Safety Lock B On Off Lever C Handpiece D Connector Plug E Thickness Control Lever ...

Page 13: ...A Screwdriver B Width Plate C Blade D Drive Pin Hole E Drive Pin ...

Page 14: ......

Page 15: ......

Page 16: ...METHOD I A Handpiece B Pocket METHOD II A Handpiece B Forceps ...

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ......

Page 21: ......

Page 22: ......

Reviews: