ziltek REMSCAN Quick Start Instructions Download Page 2

SCANNING SAMPLES

13.  Prepare soil scanning area.

a.  Remove any gross pieces of organic matter (e.g. bark, woodchip, straw) or foreign objects 

(e.g. brick, stones, gravel, glass) from the soil surface to be scanned.

b.  Firmly  compress  the  soil  surface  to  be  scanned,  so  the  surface  is  flat  and  firm  using  the

 

tamping tool provided in the kit (or a hand trowel).

c.  If the soil surface to be scanned contains excessive moisture, air-dry the sample before 

scanning. The TPH prediction is likely to be more accurate if the soil is dryer.

14.  Enter the Sample ID into the PDA (up to 20 characters, please note “\”, “/” and “:” are not allowed).

15.  Place the nose cone of the instrument in direct contact with the soil ensuring that the nose cone 

face is flush against the soil surface. Use the weight of the instrument to hold the nose cone in 

place – DO NOT push down on the instrument as this may cause a reading failure due to soil being 

forced into the nose cone.

16.  Press  the  trigger  button  once  and  release.  A  blue/green  bar  will  appear  to  progress  over  15

 

seconds and a TPH reading in mg/kg (C

10

-C

36/40

) will be displayed on the screen. You MUST hold 

the instrument as still as possible during the scanning period.

 

If RemScan gives a reading greater than 1,000, clean the nose cone prior to next measurement to 

avoid cross contamination onto the next sample.

17.  To scan more soil samples repeat steps 12 to 15 above and follow the prompts on the screen.

DOWNLOADING DATA AND SWITCHING OFF

18.  To switch off the PDA, firstly press the 

“RemScan Program Exit Button” 

on the screen. Once at 

the home page, the Bluetooth connectivity automatically switches off then the screen should then 

switch off. 

19.  Once the PDA is switched off, turn the RemScan off by pressing the power button on the bottom 

end of the handle once and release. The green light in the centre of the power button will turn red 

and then switch off.

20.  To download scanning data, remove the SD card from the slot on the top edge of the PDA by 

removing the rubber cover and pushing the spring loaded SD card downwards until it pops out. 

Insert into a computer, click on the SD card folder and download the data log in the “current date.

CSV” file from the “UserData” folder. This can also be accessed via USB connection. REMEMBER 

to reinsert the SD card before resuming scanning. 

On/off button

Background and 

Reference scan button

Timer count-down for 

Background and  

Reference scans

TPH reading mg/kg

Scan progress bar

RemScan screen 

Admin button

Scan “Trigger” button

Bluetooth 

communications

RemScan “START”

Keyboard 

raise/lower

RemScan program 

“Exit” button

Sample ID and number

Switch lock 

“slide”

Battery charge indicators

PDA - RemScan

Calibration Model

type selected

Bluetooth status indicator

Indicative 

moisture reading

Power Supply

Sylus pen access

Reset button 

* Note: Reset button is located on 

the back of the 655 PDA model

Reviews: