MU1001K
12370-0707-01
Seite
11 / 12
www.ziehl.de
9
Technical data
Rated supply voltage Us
AC/DC 24
– 240 V
0/50/60Hz
Tolerance
DC 20,4 - 297 V
AC 20 - 264 V
Power consumption
< 3 W
< 8 VA
Input
Input-
resistance
Maximum
Input signal
Error of full
scale
DC 300 V
500 kΩ
DC ± 300 V
0,1 %
DC 10 V
500 kΩ
DC ± 300 V
0,1 %
DC 60mV
10 MΩ
DC ± 2 V
0,1 %
DC 150mV
10 MΩ
DC ± 2 V
0,1 %
DC 300mV
10 MΩ
DC ± 2 V
0,1 %
Resolution
14 Bit
Measuring time
< 20 ms
Output
2 outputs with common ground
Response-time T09
< 40 ms
Voltage output
DC 0/2
– 10 V
Tolerance
0,3 % of full scale (from 0,1 V)
Temperature factor
< 0,01 % / K
Resolution
11,6 Bit
< 3,1 mV
Load
≥ 1 k Ω
Current output
DC 0/4
– 20 mA
Tolerance
0,3 % of full scale (from 0,1 mA)
Temperature factor
< 0,015 % / K
Resolution
11,6 Bit
< 6,1 µA
Load
≤ 500 Ω
Error from load
(250 Ω – resistance) / 250 Ω * 0,3 % of final value
Galvanic insulation
Us
– input - output
Test voltage
Us
– output
DC 3540V
Us
– input
DC 3540V
Input
– output
DC 3540V
Test conditions
EN 61010-1
Rated supply voltage Us (terminals A1,A2)
Pollution degree
2
Overvoltage category
III reinforced insulation
Rated insulation voltage Ui
300 V
Input (terminals 5,6,7)
Pollution degree
2
Overvoltage category
II reinforced insulation
Rated insulation voltage Ui
300 V
EMC-tests
EN 61326-1
Equipment intended for use in industrial locations
Emission
EN 61326-1, CISPR 11 class B
Immunity
EN 61326-1 industrial environment