13 Maintenance

Attention !

Disconnect the instrument from the measuring circuit before opening
the instrument to replace the battery or the fuse !

13.1  Battery

Before initial Start-up, or after long storage of your instrument, make
sure that no leakage has occurred at the instrument battery. Repeat this
inspection at regular intervals.

Battery symbol “

” appears on the LCD in either of below conditions

:
1. Battery voltage is <5.4 V. Battery replacement is necessary.
2. For a specific range depending on application

In either of above condition when the battery symbol is displayed, no
measurement can be done, and the battery symbol flashes.

Replacing the Batteries

*

Lay the instrument on to its face, loosen the two screws at the rare
panel and lift out the housing base, starting at the bottom (a).The
housing base and the housing top are held together with snap hooks
at the top of the front panel.

*

Remove all the cells from the battery holder and replace them with
new ones.

*

Important for reassembly : first lay the housing base into place such
as shown in diagram below. Push the housing base and housing
top together, first at the bottom front (a) and then at the top front (b).

*

Retighten the housing base with the two screws.

*

Please dispose of the used cells properly.

13.2  Fuses

Application of high energy pulse during Insulation resistance
measurement or application of voltage during Continuity and Ohm
measurement will blow the fuse. If the fuse blows, eliminate the cause of
the overload before placing the instrument back in to operation !

Fuse Testing

*  Set the function selector switch to any insulation measurement range.
*  Open circuit the test leads.
*  Press the “Test” button until “OL” or “FUSE” appears on display.
If display = “OL” : Fuse is OK.

If display = “FUSE” : Fuse is blown and needs replacement.

Note :

Once “FUSE” is displayed all the functions are inhibited.

Immediately switch the Instrument OFF and replace the fuse.

(b)

(a)

www.ziegler-instruments.com

10

| 10

Replacement of Fuses

*  Open the instrument as described under battery replacement.
*  Remove the defective fuse and replace it with a new fuse.
*  Make sure that the new fuse makes good contact.

Specification :

Current Rating : 500 mA / 500V-6.3 mm x 32 mm with a breaking capacity
of 10kA.
Recommended fuse manufacturer : Siba
Part No. : 70-065-63/0.5A

Be absolutely sure that only the specified fuses are used !
The use of a fuse with different triggering characteristics,
a different nominal current or a different breaking capacity
exposes the operator, the system and the measuring
instrument to danger. The use of repaired fuses or short-
circuiting of the fuse holder is prohibited.

13.3 Housing

No special maintenance is required for the housing. Excessive
contamination has an adverse effect on isolation and reduces input
resistance.The surface must be kept clean for this reason. Use a slightly
dampened cloth for cleaning. Oil free petrol is recommended for
cleaning. Avoid the use cleansers, abrasives or solvents.

ZIEGLER INSTRUMENTS

Schnepfenreuther Weg 6, D-90425 Nürnberg, Germany.

(+49)(911) 38 492 45

T E L .

(+49)(911) 32 26 212

FA X .

[email protected]

www.ziegler-instruments.com

E-MAIL

WEBSITE

in
Germany

made

Summary of Contents for RI 10

Page 1: ...15030858 www ziegler instruments com 1 10 Ziegler Instruments Order No E1 R0 921022 2 2014 EN RI 10 User Manual RI 10 Analog Digital Insulation Tester User Manual Analogue Meters With Moving Iron Movement ...

Page 2: ...On Off push button selected functions 3 Push button 11 Display for selected function 4 Push button 12 Display for the unit of 5 TEST Push button measured quantity 6 Function selector switch 13 Over range Indication 7 Terminal sockets 14 Pointer for Analog indication 8 Symbol for 15 Scale for Analog Indication CONTINUOUSLY ON 16 Activated stop watch Indicator 9 Digital display with indication 17 Ze...

Page 3: ...CE EU Confirmity mark Repair Parts Replacement and Calibration After opening the instrument live parts may be exposed Therefore the instrument must be disconnected from the measuring circuit prior to opening its case for repair replacement of parts or calibration If repair or calibration cannot be avoided unless the instrument is open and live the work must be performed by a qualified person who u...

Page 4: ...t probes after insulation resistance measurement has been activated at the instrument A current with a value of 2 mA limited by the instrument may flow over your body and although this is not life endangering the electric shock is distinctly perceptible If you are taking a measurement at a capacitive device under test for example a cable it may be charged with as much as 1000V depending upon the s...

Page 5: ...each range of insulation resistance measurement as well as for Ohms Continuity ranges Refer to the 11 Menu structure section for how to set the value for the noGo value 6 2 Storing the measured resistance value l To store the value of resistance press STORE key when the resistance measurement is in progress This will store the resistance measured at the instant the key is pressed l This way maximu...

Page 6: ...d resistance value will be displayed on the display The instrument generates a continuous sound signal if the measured resistance value is less than the pre setted GoNoGO value Default value is 0W Refer to the 11 Menu structure section for how to set the value for the limit value 7 3 Zero adjustment When measuring very low resistance values you can eliminate resistance of leads and contact resista...

Page 7: ... and Continuity Tests are inhibited When external voltage 25 V is present only the Insulation Tests are available Note The default live circuit warning does not operate i when in STOPWATCH mode ii when in MENU structure RI 10 M W 50V 100V 250V 500V 1KV V www ziegler instruments com 7 10 10 Stop Watch This function allows you to measure elapsed time up to one hour To use stop watch 1 Set the functi...

Page 8: ...Table for default values NoGo ti reCI 50V 100V 250V 500V 1000V Ohm Cont V 1 MW 1 MW 5 MW 5 MW 500 MW 00 W 5 W 00 00 Manual 0 00 0 00 0 00 0 00 0 00 0 00 0 00 www ziegler instruments com 8 10 12 Specifications Meas Range Resolu Accuracy Overload function tion of rdg value and digit duration 1 For insulation resistance range 2 For low Ohms continuity Ranges Terminal voltage on open Open circuit volt...

Page 9: ...ent which may remain unused for extended period of time Auto turn OFF Instrument turns off automatically if any of the keys or the function selector switch have not been activated for about 10 minutes in insulation resistance range and 5 minutes in other ranges Cable Set 2 Cu cross sectional area 1 mm End metallic parts should pass terminal 4 D 11 Technical data as per IEC 61010 EN 61010 2 031 CAT...

Page 10: ...he overload before placing the instrument back in to operation Fuse Testing Set the function selector switch to any insulation measurement range Open circuit the test leads Press the Test button until OL or FUSE appears on display If display OL Fuse is OK If display FUSE Fuse is blown and needs replacement Note Once FUSE is displayed all the functions are inhibited Immediately switch the Instrumen...

Reviews: