3.4
ENTERING DATA
•
Programmed start.
The oven can be programmed to come on at a set time even when no members of personnel
are in the laboratory ( see paragraph 7.3.1).
The time interval is expressed in:
- hours (0:99)
- minutes (0:59).
Use of the delayed start feature is not, of course, necessary for proper oven operation.
• Entering temperature data.
Temperature, expressed in degrees centigrade (°C) may vary within the following range:
- minimum (room temperature)
- maximum (1,100°C).
• Entering data for fast temperature increase rate.
The temperature increase rate, expressed in degrees centigrade (°C) per minute may vary within
the following range:
- minimum (1°C per minute)
- maximum (maximum power of each series D oven- see paragraph 4.1.1).
After entering the temperature increase rate parameter it is also possible
to enter the aspirator
fan switch-off temperature
(see para. 7.3.1 point I.).
1.
led HEATING
Shows, on display, that element is on
2.
led TIMING
Shows, on display, that delayed start is on
3.
led LOAD ON
Shows, on display, that active heating phase is on
4.
Data
display
5.
led °C / MIN.
Shows temperature increase rate setting on display
6.
led °C
Shows temperature setting on display
7.
Main
switch
8.
TEMP.
Shows in-chamber temperature
9.
CYCLE
Standard heating cycle with delayed start disabled
10.
DELAY CYCLE
Delayed start cycle and subsequent heating phase
11.
PROG.
Access to programmes
12.
+
To increase data settings (Time, Temperature, Temp. increase rate)
13.
_
To decrease data settings (Time, Temperature, Temp. increase rate)
3.3.1 Description of control panel components