background image

© Copyright 2004 ZEVEX Incorporated

41

WHO TO CALL

Additional Information:

YOUR HEALTHCARE PROVIDER:
_______________________________________________________________

_______________________________________________________________

_______________________________________________________________

_______________________________________________________________

_______________________________________________________________

YOUR PHYSICIAN:
_______________________________________________________________

_______________________________________________________________

_______________________________________________________________

_______________________________________________________________

_______________________________________________________________

ZEVEX Incorporated
4314 ZEVEX Park Lane
Salt Lake City, Utah 84123 USA
Customer Service and Technical Support: (800) 970-2337
www.zevex.com

vi

EnteraLite

®

Infinity

Operator’s Manual

25356_pages.qxd  2/2/2005  2:49 PM  Page 6

To Purchase, Visit 

Avobus.com

 or call 

1-800-674-3655

Summary of Contents for EnteraLite Infinity

Page 1: ...www zevex com 2004 ZEVEX Incorporated U S Patent No s 5531680 5733061 6523414 6531708 6659976 6623447 6636010 6750468 Other U S and foreign patents pending Made in USA 25356 001 Rev 01 OPERATOR S MAN...

Page 2: ...Priming and Loading the Disposable Set 7 Operating the Pump A Single Feeding Example 10 An Interval Feeding Example 13 The Battery Running on Battery Power 17 Charging the Battery 18 User Preference S...

Page 3: ...ensor 1 Keypad 1 3 Pump Instruction Label 2 Pump Wheel 1 Receptacle for Cassette 1 Receptacle for Pole Clamp 2 Speaker 2 Upstream Pressure Sensor 1 R RATE DOSE Key 3 Rinse 28 RUN PAUSE Key 3 S Self Te...

Page 4: ...port physical activity iii Index F Feed Int Key 3 Feed Interval 3 13 25 26 32 Feeding Completion 12 16 Feeding Example Single Feeding Example 10 Interval Feeding Example 13 Flow Rate 10 14 31 34 Foam...

Page 5: ...or when pump is on Do not use EnteraLite Infinity Enteral Feeding Pump for delivery of non enteral solutions Serious injury may result Proper operation of pump requires door is closed and latched Mak...

Page 6: ...val Feeding Pump enters pause mode or turns off but not due to low battery CAUTIONS Federal law U S A restricts this device to sale by or on the order of a physician registered dietitian registered nu...

Page 7: ...______ YOUR PHYSICIAN _______________________________________________________________ _______________________________________________________________ __________________________________________________...

Page 8: ...RZ Alarm 22 3 2 LOAD SET Alarm 22 3 3 LOW BATT Alarm 22 3 4 NO FLOW IN Alarm 23 3 5 NO FLOW OUT Alarm 23 3 6 NO FOOD Alarm 24 3 7 PUSH RUN TO FEED Alarm 25 3 8 DOSE DONE Message 25 3 9 NEXT DOSE Messa...

Page 9: ...1 Program Dose to 500 ml hr 10 2 12 Clear Feed Interval 11 2 13 Pump is Running 11 2 14 Amount Currently Delivered 11 2 15 Total Amount Delivered 11 2 16 DOSE DONE 12 2 17 Program Rate to 50 ml hr 14...

Page 10: ...lue rapidly Press key to decrease displayed feed function Press and hold key to decrease value rapidly Note The following keys only function when pump is in pause mode PRIME CLEAR and Figure 1 3 Keypa...

Page 11: ...l buyer ZEVEX INCORPORATED ZEVEX warrants all new EnteraLite Infinity products of its manufacture to be free from defects in material and workmanship and will replace or repair F O B at its factory in...

Page 12: ...gned in accordance with EN 60601 1 2 EN 60601 1 4 and RTCA DO 160D standards for electromagnetic emissions and immunity Normal operation should not interfere with nor be affected by other electronic e...

Page 13: ...tering the enclosure Water jets from any direction shall have no effect This symbol is printed on pump delivery sets It indicates only one patient should use each disposable set This symbol is printed...

Page 14: ...3 Figure 2 1 Spike Container Figure 2 2 Fill Bag Figure 2 3 Remove Cover from Barbed Adapter Place Cover on Pole Clamp 7 IMPORTANT PEDIATRIC CONSIDERATIONS 7 Important Pediatric Considerations Entera...

Page 15: ...the tubing below the symbol to avoid damaging the in line occluder NOTE Air may also be removed from tubing using the pump prime feature See step 7 for instructions on using the pump prime feature Ste...

Page 16: ...of 600 ml hr Display will read TO STOP LET GO Figure 2 9 Once all air is removed from tubing release key Pump will stop display will revert to last programmed rate and pump will be in pause mode For p...

Page 17: ...key to change rapidly Figure 2 10 Step 5 Press the RATE DOSE key to display dose Press the or key until a dose of 500 ml is displayed Figure 2 11 2 Directions for Use 10 EnteraLite Infinity Operator s...

Page 18: ...nt This counter never resets itself It can only be reset by the user Figure 2 15 While pump is running the settings cannot be changed and the Prime fea ture is disabled If you want to stop pump at any...

Page 19: ...he end of the feeding beep inter mittently and display will read DOSE DONE Figure 2 16 To silence alarm when the dose is complete press and hold the ON OFF key for 1 5 seconds to turn pump off or pres...

Page 20: ...rate 13 Key Sequence for An Interval Feeding Example 50 ml h 100 ml 6 00 hr 4 CLEANING 4 Cleaning WARNING To avoid electrical shock never clean pump or EnteraLite Infinity AC Adapter Charger with cha...

Page 21: ...ient s enteral feeding tube Copyright 2004 ZEVEX Incorporated 27 Figure 2 17 Program Rate to 50 ml hr Figure 2 18 Program Dose to 100 ml Figure 2 19 Program Feed Interval 6 00 hr 3 Alarms Messages and...

Page 22: ...Delivered Figure 2 22 Total Amount Delivered 3 Alarms Messages and Indications Indications CHEK Figure 3 10 Pump will sound a dual tone beep once and CHEK RATE DOSE and FEED INTERVAL will all blink on...

Page 23: ...004 ZEVEX Incorporated 25 Figure 2 23 Time Until Next Feeding 3 Alarms Messages and Indications PUSH RUN TO FEED Figure 3 7 Pump will sound a dual tone beep repeatedly and dis play will read PUSH RUN...

Page 24: ...rms Messages and Indications NO FOOD Figure 3 6 Pump will sound a dual tone beep repeatedly and dis play will read NO FOOD Why This alarm has occurred because pump has detected air in the tubing It ta...

Page 25: ...blockage in the set between pump and bag To correct the problem Press the RUN PAUSE key to silence alarm and place pump in pause mode then check the following 1 Check tubing for kinks or formula clum...

Page 26: ...e Settings BEEP LOW and BEEP HIGH Figure 2 27 Settings Lock UNLK and LOCK Figure 2 28 Dose Done Settings BEEP WHEN DONE and MUTE WHEN DONE Figure 2 29 Backlight Settings LITE OFF and LITE ON 3 Alarms...

Page 27: ...seconds the and the ON OFF keys at the same time Pump will beep and then display the words BEEP HIGH or BEEP LOW Press the key to change to HIGH or the key to change to LOW Pump will give a triple bee...

Page 28: ...keys at the same time Pump will beep and then display the words BEEP HIGH or BEEP LOW Press the key to change to HIGH or the key to change to LOW Pump will give a triple beep when the or keys are pre...

Page 29: ...e Settings BEEP LOW and BEEP HIGH Figure 2 27 Settings Lock UNLK and LOCK Figure 2 28 Dose Done Settings BEEP WHEN DONE and MUTE WHEN DONE Figure 2 29 Backlight Settings LITE OFF and LITE ON 3 Alarms...

Page 30: ...because pump has detect ed blockage in the set between pump and bag To correct the problem Press the RUN PAUSE key to silence alarm and place pump in pause mode then check the following 1 Check tubing...

Page 31: ...ns NO FOOD Figure 3 6 Pump will sound a dual tone beep repeatedly and dis play will read NO FOOD Why This alarm has occurred because pump has detected air in the tubing It takes approximately 1 ml of...

Page 32: ...004 ZEVEX Incorporated 25 Figure 2 23 Time Until Next Feeding 3 Alarms Messages and Indications PUSH RUN TO FEED Figure 3 7 Pump will sound a dual tone beep repeatedly and dis play will read PUSH RUN...

Page 33: ...ure 2 21 Amount Currently Delivered Figure 2 22 Total Amount Delivered 3 Alarms Messages and Indications Indications CHEK Figure 3 10 Pump will sound a dual tone beep once and CHEK RATE DOSE and FEED...

Page 34: ...e Copyright 2004 ZEVEX Incorporated 27 Figure 2 17 Program Rate to 50 ml hr Figure 2 18 Program Dose to 100 ml Figure 2 19 Program Feed Interval 6 00 hr 3 Alarms Messages and Indications Battery Doesn...

Page 35: ...rate 13 Key Sequence for An Interval Feeding Example 50 ml h 100 ml 6 00 hr 4 CLEANING 4 Cleaning WARNING To avoid electrical shock never clean pump or EnteraLite Infinity AC Adapter Charger with cha...

Page 36: ...NE the pump will stop at the end of the feeding beep inter mittently and display will read DOSE DONE Figure 2 16 To silence alarm when the dose is complete press and hold the ON OFF key for 1 5 second...

Page 37: ...ts itself It can only be reset by the user Figure 2 15 While pump is running the settings cannot be changed and the Prime fea ture is disabled If you want to stop pump at any time press the RUN PAUSE...

Page 38: ...key to change rapidly Figure 2 10 Step 5 Press the RATE DOSE key to display dose Press the or key until a dose of 500 ml is displayed Figure 2 11 2 Directions for Use 10 EnteraLite Infinity Operator s...

Page 39: ...mping at the maximum rate of 600 ml hr Display will read TO STOP LET GO Figure 2 9 Once all air is removed from tubing release key Pump will stop display will revert to last programmed rate and pump w...

Page 40: ...l to avoid damaging the in line occluder NOTE Air may also be removed from tubing using the pump prime feature See step 7 for instructions on using the pump prime feature Step 5 Loop silicone tubing a...

Page 41: ...3 Figure 2 1 Spike Container Figure 2 2 Fill Bag Figure 2 3 Remove Cover from Barbed Adapter Place Cover on Pole Clamp 7 IMPORTANT PEDIATRIC CONSIDERATIONS 7 Important Pediatric Considerations Entera...

Page 42: ...rotection against water entering the enclosure Water jets from any direction shall have no effect This symbol is printed on pump delivery sets It indicates only one patient should use each disposable...

Page 43: ...60601 1 2 EN 60601 1 4 and RTCA DO 160D standards for electromagnetic emissions and immunity Normal operation should not interfere with nor be affected by other electronic equipment EnteraLite Infini...

Page 44: ...l buyer ZEVEX INCORPORATED ZEVEX warrants all new EnteraLite Infinity products of its manufacture to be free from defects in material and workmanship and will replace or repair F O B at its factory in...

Page 45: ...nd hold key to increase value rapidly Press key to decrease displayed feed function Press and hold key to decrease value rapidly Note The following keys only function when pump is in pause mode PRIME...

Page 46: ...r 10 2 12 Clear Feed Interval 11 2 13 Pump is Running 11 2 14 Amount Currently Delivered 11 2 15 Total Amount Delivered 11 2 16 DOSE DONE 12 2 17 Program Rate to 50 ml hr 14 2 18 Program Dose to 100 m...

Page 47: ...RZ Alarm 22 3 2 LOAD SET Alarm 22 3 3 LOW BATT Alarm 22 3 4 NO FLOW IN Alarm 23 3 5 NO FLOW OUT Alarm 23 3 6 NO FOOD Alarm 24 3 7 PUSH RUN TO FEED Alarm 25 3 8 DOSE DONE Message 25 3 9 NEXT DOSE Messa...

Page 48: ...________________________________ YOUR PHYSICIAN _______________________________________________________________ _______________________________________________________________ ________________________...

Page 49: ...use mode or turns off but not due to low battery CAUTIONS Federal law U S A restricts this device to sale by or on the order of a physician registered dietitian registered nurse or other licensed prac...

Page 50: ...or when pump is on Do not use EnteraLite Infinity Enteral Feeding Pump for delivery of non enteral solutions Serious injury may result Proper operation of pump requires door is closed and latched Mak...

Page 51: ...ortability promote and sup port physical activity iii Index F Feed Int Key 3 Feed Interval 3 13 25 26 32 Feeding Completion 12 16 Feeding Example Single Feeding Example 10 Interval Feeding Example 13...

Page 52: ...struction Label 2 Pump Wheel 1 Receptacle for Cassette 1 Receptacle for Pole Clamp 2 Speaker 2 Upstream Pressure Sensor 1 R RATE DOSE Key 3 Rinse 28 RUN PAUSE Key 3 S Self Test 9 Serial Number 2 9 Set...

Page 53: ...Priming and Loading the Disposable Set 7 Operating the Pump A Single Feeding Example 10 An Interval Feeding Example 13 The Battery Running on Battery Power 17 Charging the Battery 18 User Preference S...

Page 54: ...www zevex com 2004 ZEVEX Incorporated U S Patent No s 5531680 5733061 6523414 6531708 6659976 6623447 6636010 6750468 Other U S and foreign patents pending Made in USA 25356 001 Rev 01 OPERATOR S MAN...

Reviews: