2.4

 

Connecting ZET 7152-N-VER.3  seismograph to PC 

The PC should have Windows OS and ZETLAB software installed. It is also necessary to 

connect ZETKEY (which is included into delivery scope) to USB port.  

ZET 7152-N-VER.3  is  connected  to  PC  via  Ethernet  interface.  In  order  to  connect  the 

ZET 7152-N-VER.3 to PC via Ethernet interface, connect the cable PatchCord UTP (Cat. 5е) to 

Ethernet port of ZET 7152-N-VER.3 and free Ethernet port of the PC. As ZET 7152-N-VER.3 is 

connected to the PC, enable “Power” key at the top panel of ZET 7152-N-VER.3. The system will 

find and install the corresponding driver.  

IP address of the PC should be in the same network with IP address of interface converter 

ZET 7176 (included into seismograph). Local network should not block UDP (multicast). 

Note: By default, interface converter has IP-address 192.168.1.76 with mask 255.255.255.0.

 

Configuring of the devices included into the scope of ZET 7152-N-VER.3 is performed in 

“Device manager” program from “Service” menu of ZETLAB panel. (

Figure 2.1

). 

 

 

Figure 2.1 ZETLAB panel 

Summary of Contents for ZET 7152-N-VER.3

Page 1: ...SEISMOGRAPH ZET 7152 N VER 3 USER MANUAL LLC ETMS ...

Page 2: ...rs included into the scope of ZET 7152 N VER 3 19 2 6 1 Configuring digital short period seismometer ZET 7156 21 2 6 2 Configuring digital accelerometer ZET 7152 N 22 2 6 3 Configuring off line recorder ZET 7173 23 2 6 4 Configuring synchronization module ZET 7175 25 3 Operation of ZET 7152 N VER 3 26 3 1 Mounting of ZET 7152 N VER 3 off line use 26 3 2 Mounting of ZET 7152 N VER 3 stationary use ...

Page 3: ...structure of ZET 7152 N VER 3 which do not affect technical characteristics and do not require introducing any amendments into the technical documentation It is necessary to observe the guidelines of the present manual and other documents supplied together with ZET 7152 N VER 3 at all stages of its operation The present user manual and product certificate are included into ZET 7152 N VER 3 deliver...

Page 4: ...d for seismic impact parameters measurements ZET 7152 N VER 3 can be used off line or within the scope of automated monitoring system for the purpose of seismic and seismological research by means of reflected and refracted waves method and converted waves method ZET 7152 N VER 3 can also be used for geotechnical studies ...

Page 5: ...max 0 2 mg ZET 7152 N 0 0001 mm s ZET 7156 Sampling rate by channels 30 125 250 500 1000 2000 Hz ZET 7152 N 1 8 16 32 64 128 Hz ZET 7156 Primary transducer control Integrated electrical excitation system of sensing elements with actuator yes Primary transducer power control yes Off line mode Off line mode operation time max 18 h Battery type lead Integrated non volatile memory volume 32 Gb Time of...

Page 6: ...ttery charger SADP 65KB 220V 19V 1 pc Pins for ground mounting for off line product version 3 pcs Mounting kit for stationary use mounting plates 1 2 set of mounting screws junctions box 1 kit ZETKEY 1 pc CD with ZETLAB software 1 pc Product certificate 1 pc User manual 1 pc it is allowed to provide a single copy of User Manual for products lot up to 10 pcs 1 The guaranteed service life is only va...

Page 7: ...Tel Fax 7 495 739 39 19 www zetlab com 7 1 4 Construction of ZET 7152 N VER 3 Components of the ZET 7152 N VER 3 are placed inside of a cylindrical package Overall view of ZET 7152 N VER 3 is shown in Figure 1 1 ...

Page 8: ... signal The measuring module in its turn performs digital processing of the analog signals and transfers digital data via CAN 2 0 interface 2 Digital accelerometer ZET 7152 N Digital accelerometer ZET 7152 N is used for measurements and conversion of acceleration into a digital signal and further transfer of the data in digital format The accelerometer consists of a vibration sensor with integrate...

Page 9: ...52 N VER 3 data is transmitted via Ethernet in real time mode Battery capacity enables measurements during 18 hours Meanings of indicators located at the lid of ZET 7152 N VER 3 are shown in Table 1 3 Table 1 3 Meanings of indicators Indicator name Indicator color Indicator state Event Battery Red On It is necessary to charge the battery Blue On Battery is charging Green On Battery is charged The ...

Page 10: ...corded data to PC via USB interface Ethernet Connection of ZET 7152 N VER 3 to industrial PC via local Ethernet network Charge Connection of charging device SADP 65KB to ZET 7152 N VER 3 Sync Connection of external aerial GPS GLONASS Trimble to the ZET 7152 N VER 3 Power key is used to switch ZET 7152 N VER 3 on off In order to switch ZET 7152 N VER 3 on press and hold Power key for at least 3 sec...

Page 11: ...ess of ZET 7152 N VER 3 in accordance with the items specified in Table 1 2 see Clause Ошибка Источник ссылки не найден Perform visual examination check ZET 7152 N VER 3 for any mechanical damage Recommendation In the case if there is a room for storing of ZET 7152 N VER 3 it is recommended to save the package If ZET 7152 N VER 3 is transported it should be placed into the package so that to avoid...

Page 12: ...rge port of ZET 7152 N VER 3 panel Connect the charging device to 220 V AC network 2 During battery charging the Battery indicator should have blue color 3 Green color of Battery indicator means that the battery charging process is complete Upon charging process completion disconnect charging device from ZET 7152 N VER 3 and 220V AC Note Battery replacement is not covered by warranty provisions an...

Page 13: ...l the software Note the following system requirements should be met for proper operation of the software Dual core processor or more Core speed over 1 6 GHz RAM over 2 Gb HD free space over 20 Gb Video controller 3D graphics acceleration support of ОреnGL DirectХ memory over 128 Mb Resolution min 1280 1024 Mouse or any other pointing device Standard keyboard or any other entry device CD ROM for pr...

Page 14: ...ET 7152 N VER 3 and free Ethernet port of the PC As ZET 7152 N VER 3 is connected to the PC enable Power key at the top panel of ZET 7152 N VER 3 The system will find and install the corresponding driver IP address of the PC should be in the same network with IP address of interface converter ZET 7176 included into seismograph Local network should not block UDP multicast Note By default interface ...

Page 15: ...erface converter activate it from the menu with right click on its serial number Figure 2 3 Figure 2 3 Connecting to the interface converter Note Serial number of interface converter is specified in its product certificate During connection process the device name will be highlighted with bold type If the device name is depicted in bold type for a long time it means that it is impossible to establ...

Page 16: ...ess identical with that of the interface converter In order to check current IP address of the interface converter enter the menu with right click on the interface converter enter Properties menu and go to Device tab Figure 2 4 The parameter IPv4 Address will indicate the current IP address of the interface converter Figure 2 4 Device tab ...

Page 17: ...interface converters and devices that are connected to PC The second level shows sensors that are connected to a particular interface converter To the right from the digital sensor one can see device address in the measuring network Figure 2 6 Figure 2 6 Addresses of the devices in the measuring network In the case if detailed display mode is enabled the right section of the window will display ba...

Page 18: ...ng ZET 7176 interface converter In order to configure interface converter ZET 7176 one should Set the network address in accordance with applicable requirements in the fields Ipv4 address Subnet mask Default gateway of Ethernet tab Figure 2 7 Figure 2 7 Ethernet tab ...

Page 19: ...gure 2 8 CAN tab Note After changing the bit rate of interface converter it is necessary to cycle power ZET 7152 N VER 3 The system will automatically change the bit rate of digital sensors which are connected to the interface converter Note additional information regarding configuration of ZET 7176 interface converter is available in the document ZET 7176 User Manual ...

Page 20: ...d by the manufacturer Please note that in the field Node 2 63 of Information tab there should be set a unique address of each device of the measuring network All the devices within the scope of the seismograph should have unique address it is an obligatory condition for normal operation of ZET 7152 N VER 3 The nodes are set in the range from 3 to 63 Figure 2 9 shows an example of Information tab o...

Page 21: ...the case if the first channel device address is changed the system will automatically change the addresses of the remaining channels by 1 Note digital accelerometer ZET 7152 N consists of three channels by default named as ZET 7152_X ZET 7152_Y ZET 7152_Z and has three addresses in the measuring network It is possible to configure the seismometer parameters only in the tab of the first channel ZET...

Page 22: ... seismometer ZET 7156 is performed in the Settings tab of Properties menu Figure 2 10 In order to configure the seismometer set corresponding parameters in Configuration tab Figure 2 10 Configuration tab Note additional information regarding configuration of short period seismometer ZET 7156 is available in ZET 7156 user manual ...

Page 23: ...rometer ZET 7152 N is performed in Settings tab of Properties menu Figure2 11 In order to configure the accelerometer set corresponding parameters in Settings tab Figure2 11 Settings tab Note additional information regarding configuration of digital accelerometer ZET 7152 N is available in the document ZET 7x52 N User Manual ...

Page 24: ...com 24 2 6 3 Configuring off line recorder ZET 7173 In order to configure off line recorder ZET 7173 it is necessary to Select property All addresses in the field Data recording mode of Recording tab Figure 2 12 Figure 2 12 ZET 7173 Recording tab ...

Page 25: ...tlab com 25 select Linear property in the field Type of record of Storage tab Figure 2 13 Figure 2 13 ZET 7173 Storage tab Note additional information regarding configuration of off line recorder ZET 7173 is available in ZET 7173 User manual ...

Page 26: ...4 Figure 2 14 ZET 7175 Configuration tab In order to configure synchronization module ZET 7175 one should set the following parameters in the Configuration tab Figure 2 14 1 Select On in the field GPS receiver activation 2 Select Off in the field LED indication for receiver status Note additional information regarding configuration of synchronization module ZET 7175 is available in the document ZE...

Page 27: ...lace ZET 7152 N VER 3 in the installation area align measuring axis X by azimuth normally direction to the North is used Push ZET 7152 N VER 3 into the ground the pins should be placed into the ground by 2 3 of their length If necessary use additional pressure to align ZET 7152 N VER 3 by horizontal plane use bubble cell to check the level 3 2 Mounting of ZET 7152 N VER 3 stationary use Mounting s...

Page 28: ...йден click Power key at the front panel of ZET 7152 N VER 3 Set the necessary parameters see Clause Ошибка Источник ссылки не найден As the parameters are set switch on ZET 7152 N VER 3 with Power key Mount ZET 7152 N VER 3 in the area of seismic research performance see Clause Ошибка Источник ссылки не найден Connect GPS aerial to Sync port of ZET 7152 N VER 3 Switch on ZET 7152 N VER 3 with Powe...

Page 29: ...its connection to local automated monitoring system for the purpose of further data transfer to server and AWS systems Configuration of server AWS and ZET 7152 N VER 3 within the scope of automated monitoring system is to be performed in accordance with Administrator manual which is included into the automated control system documentation Operator s manual describes guidelines regarding operation ...

Page 30: ...s to PC memory ZETLAB software uses two directories one for saving signals and another one for storage of compressed signals data In order to set directories configuration it is necessary to activate ZETLAB icon and enable the panel User s path configuration in the window Main menu of the control panel Figure 3 1 Figure 3 1 Main menu control panel ...

Page 31: ...n access Figure 3 2 enable the panel for each of the directories corresponding to the type of the saved data signals compressed signals you will see the window Choose the directory Here you can assign configuration path and enable Apply key Figure 3 2 Window Adjusting configuration access ...

Page 32: ...process of files saving and conversion into the file directories for storing signals and compressed signals data Figure 3 2 5 Upon completion of files copying and conversion you will see a window offering you to start View historical events program see ZETLAB Software Operator s Manual Note In the case if Cancel key of Select files to convert window is enabled Figure 3 3 the corresponding window w...

Page 33: ...SB interface Make sure that there is a direct USB connection between PC and ZET 7152 N VER 3 without any intermediate devices like HUB etc Connection failure of ZET 7152 N VER 3 via Ethernet interface Make sure that ZET 7152 N VER 3 and the PC are located in one and the same subnet see Clause Ошибка Источник ссылки не найден ...

Page 34: ... com 34 5 Technical maintenance Technical maintenance of ZET 7152 N VER 3 should be performed at least once in 6 months integrated battery should be charged in accordance with the requirements specified in Clause Ошибка Источник ссылки не найден ...

Page 35: ...tion process Transporting allocation of ZET 7152 N VER 3 not more than in two rows Transportation climatic requirements Ambient temperature from 35 up to 60 С Relative humidity up to 98 at the temperature of 25 С Atmospheric pressure from 84 up to 107 kPa from 630 up to 800 mmHg In the course of transportation process it is necessary to protect ZET 7152 N VER 3 package from atmospheric precipitati...

Page 36: ...elines for mounting plate selection In the case of mounting at horizontal surface one should use Plate 1 Wafer 2 is used in the case of horizontal mounting when adjustment is required rotation of ZET 7152 N VER 3 by azimuth 360 degrees and angle 3 degrees For mounting at vertical surface one should use Plate 1 and Bracket U piece and a plate with a thickness over 6 mm welded to it not included int...

Page 37: ...LLC ETMS Tel Fax 7 495 739 39 19 www zetlab com 37 Connection dimensions of Mounting Plate 1 ...

Page 38: ...495 739 39 19 www zetlab com 38 Record of changes Ame ndme nt Number of sheet s page s Total number of sheet s page s Docum ent numbe r File and date of accompanyin g document Signatur e Date Amended Substitute d New Void ...

Reviews: