All components and instruments are supplied NON-STERILE. 

 

Check instruments for damaged surfaces, chipped off edges, deformations and eliminate 

damaged instruments. Products that are still contaminated need to be cleaned and disinfected 

until all visible contaminants have been removed. 

 

Arrange the cleaned and disinfected instruments in the dedicated bur block. Wrap the bur block 

using disposable sterilization packaging and/or sterilization containers. Sterilize the assembled 

bur block according to the parameters listed in the following table: 

 

 

Gravity Autoclave 

Pre-Vacuum Autoclave 

Temperature 

132°C 

132°C 

Exposure Time 

25 minutes 

4 minutes 

Dry Time 

30 minutes 

20 minutes 

 

 

WARNINGS AND PRECAUTIONS  

 

The use of this or any dental restorative product requires that the clinician and other specialists be 

thoroughly familiar with the use and application of these instruments. They must also be familiar 

with the instruments and kit sterilization requirements. 

 

Care should be taken in choosing the operating speed and force applied to instrument during use to 

avoid damaging the instrument or work piece due to excessive heat generation. 

 

Improper use of instruments may contribute to premature wear of the instruments, lead to poor 

results and pose a danger to the operator. 

 

The Polisher is for single use only in cases where residual blood and/or saliva are present. 

 

The Kit and individual instruments are packaged NON-STERILE and MUST BE properly cleaned, 

disinfected, and sterilized per instructions below prior to use. 

 

Refer and strictly adhere to the manufacturers’ instructions for use for the reaction time and 

suitability of disinfectants and detergents. 

 

DIRECTIONS FOR USE - DENTURE PREPARATION PROCEDURES

 

The steps below are guidelines to help process the dentures for implant-retained restorations: 

1.

 

Use the Recess Bur to create a void in the denture. Use the indicators on the bur as shown below 

to create the proper recess depth for the LOCATOR and SATURNO Cap. 

 

2.

 

It is recommended that you use the Undercut Bur to create an undercut in the recess which 

enhances the mechanical retention of the pick-up material to the prosthesis. The Undercut Bur 

can also be used to create additional clearance in the recess, if needed. 

3.

 

Use the Vent Bur to create a small hole through the denture recess, which enables the visibility of 

the Denture Cap and allows for excess pick-up material to flow out. 

4.

 

Use the Trim Bur to carefully trim the flanges and any rough edges on the denture, if needed. 

5.

 

Use the Grind Bur to carefully trim around the attachment pick-up material, if needed. 