Instructions for Use 

Zest CHAIRSIDE® Denture Prep and Polish Kit 

IMPORTANT: 

This document contains the most current Instructions for Use. Please read and retain. 

 

PRODUCT DESCRIPTION 

The Zest CHAIRSIDE Denture Prep and Polish Kit contains the necessary instruments to prepare and 

finish an overdenture for pickup of a LOCATOR® or SATURNO® attachment. The system includes a recess 

bur, undercut bur, vent hole bur, trim bur, grind bur, and polisher. 
 

INDICATIONS FOR USE

 

The Zest CHAIRSIDE Denture Prep and Polish Kit is intended to prepare and polish full and partial acrylic 

dentures as part of a chairside or laboratory procedure. 
 

CAUTION

 

Federal (USA) law restricts this device to sale by or on the order of a licensed dentist. 
 

MULTI-USE DEVICES 

CHAIRSIDE Denture Prep and Polish Kit are multi-use devices and are provided non-sterile.  Reusable 

tools and instruments must be cleaned and sterilized prior to use or reuse on patients.  Reuse of a multi-

use devices without cleaning and sterilization may cause harm to the patient in the transfer of blood, 

tissue, or saliva that may contain infectious disease. 

PREPARATION

 

 

Impurities need to be removed from the products directly after use (within two hours 

maximum). To do so, use running water or an FDA approved and/or CE Marked disinfectant 

solution. The disinfectant’s effectiveness should be established, suitable, and compatible with 

the instruments. 

CLEANING INSTRUCTIONS

 

 

Place the instruments in the cleaning bath for the prescribed contact time and concentration, 

take care that the instruments do not contact each other. 

 

In case of severe contamination of the instrument, it is recommended to use an ultrasonic bath 

for cleaning. 

 

The Bur Block must be cleaned and disinfected without instruments loaded. 

 

The Bur Block is not suitable for cleaning using a thermo disinfector (disinfector RDG). 

 

Remove instruments from the cleaning solution and rinse for at least 3 minutes with tap water. 

 

Remove excess moisture from the components with a clean, absorbent, and non-shedding wipe. 

 

RECOMMENDED STERILIZATION INSTRUCTIONS

 