22 - D2O VCP Product Manual
COMMAND SYNTAX
DETAILS
sys -c
Displays the unique Network Module Registration Code. Typically the module is already
pre-registered on the network.
Example:
sys -c
tpsez74fgajae334yddtr66my
COMMAND SYNTAX
DETAILS
report
Displays a Commissioning Report of all key parameters currently set on the D2O.
IMPORTANT NOTE: The
apply
command should be used prior to executing this
command to ensure that all
Admin Restricted
parameters of
acq
and
inp
commands
have been applied, and that the Commission Report is showing correct information.
Unapplied changes will still display in the Commission Report but will not execute until
the
apply
command is used.
The returned date and time formats are: Day - Month - Year Hour : Minute : Second in
Greenwich Mean Time (GMT).
Example: Report with example data for a D2O setup in Analogue mode using a 10V input type.
report
COMMISSIONING REPORT
Date/Time: 13-01-2021 23:17:52
Model: VCP
Module ID: 007afe3da9e M1-24
FW/HW Build: FW: 1.2.4.2 HW: CUSTOM1
TX/RX Build: 1.3.2
Temp: 25.1 DegC
POWER
Batt. Charge: 90 %
Battery Lvl: 3128 mV
GPS
Latitude: -27.5775267
Longitude: 153.0969805
Acquired: 13-01-2021 23:13:01
ACQUISITIONS
Acq.Mode: Analogue
Acq. Type: Raw
Sync Time: 00:00
Data Sets Acq: 2136
Next Acq: 14-01-2021 00:00:00
Pre-Scaler: 10
Interval: 1h
Value Type: Raw
Output Off at: 80
Output On at: 50
Output State : OFF
TRANSMISSION
Messages/24h: 3 Required
Messages/24h: 3.00 Scheduled
Message ID: 24
READINGS
Input Type: 10V
Smooth Level: 1
Offset Comp.: 1200 uV/uA
Voltage Input: 3796 mV
Pulse Counts: 2150