Fixed Hoist Test Facility Operation and Maintenance Manual
ZOM-12000-7 rev 0
Page 16 of 24
This manual contains proprietary information and is not to be copied or disclosed without written
permission from Zephyr International LLC. Copyright 2006-2008
ZGS-12216-1
Load Cell Calibration
Install a hoist and make all electrical and mechanical attachments.
Configure the load cell readout according to section 4 of the DP25B-S User manual.
Input type is 100 mV
Set the decimal point to the default 0000
Dip switch setting is
12345678
COCOOCOO
Set Rd.CF R.2=1
If a 600 lbs load is available Scale the meter according to section 4.3.1.
With the UUT mounted on the hanging hoist mount with no load on it and with the
mount attached to the load cells, store the IN 1 value as the unscaled display reading
shown and set the RD 1 value to zero.
Then lower the UUT hook and attach the load and raise the 600 lbs weight until the
weight is off the ground.
Store the IN 2 reading and input the Rd 2 value as 600 lbs.
Note: This procedure can be performed with a lower or a higher load, and the Rd 2
value is then adjusted to the exact weight used. Using a value of 600 lbs insures
maximum accuracy, but is not essential.