6.3 How to generate delimiters
Connect the decoder output of the external additional circuit to DATACR, DATALF,
DATAEND according to the code of delimiter.
GP-IB DATA
Last data
LF
CR
1)
When outputting EOI at the same time as CR, LF code and LF, connect the strobe before
the last data of data strobe created by the decoder to DATACR and the strobe of the last
data to DATALF and DATAEND. It is able to be sent to up to 254 bytes of data to GP-IB.
LF
GP-IB data
Last data
CR
EOI
2)
In case of EOI delimiter.
Connect to the DATAEND of the last data of the data strobe created by the decoder.
It is able to be sent up to 256 bytes of data to GP-IB.
GP-IB data
Last data
EOI