background image

 
 

Plateforme de Microscopie Photonique 

 
 

  

Page 2 of 2 

Updated on 2019-02-12 

User Guide Zeiss Z2 B-301

 

Setup 

On the microscope touch screen: 

9.

 

Press Home>

Load Position

 

10.

 

Press 

Set Work Position

 

11.

 

Move the focus slightly upward to remove the « 

Lower Z limit reached

 » message 

12.

 

Press Home>Microscope>Turret>

Objectives

>

10x

 

13.

 

Press 

Done

 (if prompted) 

14.

 

Press Home>Microscope>XYZ>Position>

Set zero Auto

 (focus calibration) 

15.

 

Press 

OK

 

16.

 

Wait until the calibration is done 

Loading samples 

On the microscope touch screen: 

1.

 

Press Home>

Load Position

 

2.

 

Press 

Set Work Position

 

On the microscope: 

3.

 

Place your sample on the microscope stage 

Focusing with air objectives 

On the microscope touch screen: 

5.

 

Press Home>Microscope>Turret>

Objectives

 

6.

 

Press 

10x, 20x 

OR

 40x

 

On the computer: 

7.

 

Open 

Zen Blue

 software 

8.

 

In the 

Locate Tab

, select 

BF Visible

 or 

Fluo Visible

 

9.

 

Adjust the focus while looking through the eyepieces (focus is reached around Z=13mm on a calibrated focus) 

10.

 

In the 

Locate Tab

, select 

Off

 

Focusing with oil objectives 

Focus first with an air objective

 then, on the microscope touch screen: 

8.

 

Press Home>Microscope>Turret>

Objectives

 

9.

 

Press 

60x, 100x 1.3 

OR

 100x 1.4

 

10.

 

Place a drop of oil on your sample 

11.

 

Press 

Done

 

In 

Zen Blue

 software: 

12.

 

In the 

Locate Tab

, select 

BF Visible

 or 

Fluo Visible

 

13.

 

Adjust the focus with the fine adjustment wheel while looking through the eyepieces (focus is reached around Z=13mm on 
a calibrated focus) 

11.

 

In the 

Locate Tab

, select 

Off

 

 
Your sample is ready for acquisition. 
 