6.1 Troubleshooting
Description of the fault
Cause of the fault
Remedy for the fault
Controller not responding
(no LEDs lighting up)
No power supply
Establish a power supply (see 3.1).
Switch on the switch on the front of the device
(see 3.2).
Power
supply
unit
defective
Exchange the power supply unit (see 4.2)
Electronics defective
Contact ZEISS Service
Controller not responding
(blue and yellow LEDs lit up)
Electronics defective
Contact ZEISS Service
WSB Wheel handwheel not
responding
(position
is
displayed)
Handwheel is inactive.
Display
is
showing
“off“
.
Press the middle button (Fig. 4/
2
) until the
display shows “on".