Initial start-up
000000-2227-967-GA-GB-281021
29
•
The device will now start in
Refraction mode
•
If necessary, tap
Back
to return to the previous page for language
selection.
Connection to the EMR/PMS system
For connecting the device to an EMR/PMS system, there are several options:
•
Connecting via the serial interface
•
File-based data exchange via a network folder
•
Data output via network
•
Provision of data to be collected via the network (web service)
Please check which of these options are supported by your EMR/PMS
system. The current device interface specifications can be found at
www.zeiss.com/essential-line/connectivity.
Using the serial interface
•
Switch off both devices and plug the RS 232 cable into the
corresponding connecting ports of the ZEISS VISUREF 150 (
7
, Fig. 5) and
the receiving device, e.g. your i.Com server, PC or your
VISUPHOR/VISUSCREEN Junction Box. Use the fixing screws provided.
•
Switch on both devices.
•
Set the configuration parameters for both devices.
•
Switch to the device settings of ZEISS VISUREF 150 and open the
Interface/Serial
, page
).
•
Align the
Baud rate
,
Number of bits for Data and Stop
,
Parity
and
Handshaking
interface parameters with the corresponding settings of
the receiving device.
•
Select the desired output format and protocol.
•
Enable the interface in the
Interface/Interface enablement
menu (see
, page
•
Exit the device settings and tap
Apply
.