Electromagnetic
compatibility
CAUTION
–
RISK
OF
ELECTROMAGNETIC
RADIATION
The
ZEISS
VISUPHOR
500
may
not
be
placed
next
to
or
stacked
together
with
other
equipment,
except
in
the
device
configurations
described
in
this
user
manual.
If
operation
close
to
or
with
other
devices
is
necessary,
the
ZEISS
VISUPHOR
500
must
be
closely
observed
to
monitor
its
proper
functioning
in
this
configuration.
CAUTION
-
GENERAL
HAZARDS
Medical
electrical
equipment
is
subject
to
special
EMC
precautions
and
must
be
installed
and
commissioned
in
accordance
with
the
EMC
instructions
included
in
the
accompanying
documentation.
Replacement
cables
may
only
be
purchased
at
Carl
Zeiss
Meditec
or
at
dealers
authorized
by
Carl
Zeiss
Meditec.
The
use
of
accessories,
all
types
of
transducers
and
cables
not
specified
in
this
user
manual
or
not
sold
by
Carl
Zeiss
Meditec
as
replacement
parts
may
result
in
higher
emissions
or
reduced
immunity
of
the
device.
Relevant
accessories:
•
Serial
connection
cable
RS232
SUB-D,
3
m,
1:1
seria
cable,
male
<->
female
•
WLAN
access
point
for
VISUSCREEN/VISUPHOR/Access
point
Lancom
L-321agn
Wireless
US
•
Power
cables
•
Bluetooth/USB
adapter
CAUTION
-
GENERAL
HAZARDS
If
the
device
shows
obvious
damage
(e.g.
on
housing
or
cables),
ZEISS
Service
should
be
informed.
It
is
possible
that
ZEISS
VISUPHOR
500
may
still
function
if
damaged,
but
is
much
more
susceptible
to
electromagnetic
radiation.
59
000000-2075-933-GA-en-GB-230822