4.3 Mounting the foot switch on the foot of the instrument table
000000-2096-101-DokS-GB-181218
45 / 116
4.3 Mounting the foot switch on the foot of the
instrument table
To fasten the TouchControl foot switch to the corrugated feet of
the IT 760.i / IT 1060.i instrument tables, you must first remove the
base plate of the foot switch. To do so, unscrew the front fastening
screws and loosen the two rear screws. The installation is
performed as described below.
The feet of the IT 760.i / IT 1060.i instrument tables have three /
five drilled holes to which the foot switch can be secured by means
of a screw. Each of the holes is covered by a grub screw.
1
2
3
4
5
6
7
8
Figure 12: Mounting the foot switch onto the instrument table
1
Drill hole in the table foot
2
Grub screw for covering drill holes in the
table foot
3
Adapter plate
4
Drill hole in the adapter plate
5
Assembly hole in the rotating plate
6
Screw for fastening the adapter plate to the
table foot
7
Foot switch
8
Screws for fastening the foot switch to the
adapter plate