Documentation set
12 Optional accessories
SL 800
12.5 Focusing rod
000000-2268-355-DokS-GB-300719
71 / 94
12.5 Focusing rod
The focusing rod can be used to compensate for individual
ametropia of the user or any device ametropia by adjusting the
diopter scale on the eyepieces. Exact adjustment of the eyepieces is
a prerequisite for optimum examination results and is necessary
when using accessories for image documentation. Otherwise, the
image may appear to be sharp during observation, but a blurred
image may be captured by the camera, see Adjusting the eyepieces
and interpupillary distance [
37].
Figure 22: Focusing rod
12.6 AT 030 applanation tonometer
The applanation tonometer is for measuring the intraocular
pressure of a seated patient. The correct tonometer holder should
be used for mounting on the various ZEISS slit lamps.
1
2
Figure 23: SL 800 with AT 030 applanation tonometer
1
Tonometer holder
2
AT 030 applanation
tonometer
Observe the instructions for use of the applanation tonometer.