7 Troubleshooting | 7.1 Chamber
ZEISS
Symptom
Cause
Measure
Stage does not move
Stage needs to be initialized
Refer to
If this does not solve the prob-
lem, contact your local ZEISS
service representative
Stored position of the
specimen stage cannot be
approached correctly
Absolute stage movement is
required, stage needs to be
driven to a well-defined posi-
tion
Refer to
Microscope does not vent
No nitrogen
Check nitrogen supply
No compressed air
Check compressed air supply
Vac ready = OK
is not
displayed after specimen
exchange
System vacuum is bad due to a
vacuum leak at the chamber
door
Check the chamber door seal
for cleanliness
Vac ready = OK
is dis-
played very late after
specimen exchange
Gas ballast is activated at the
pre-vacuum pump
Deactivate gas ballast at the
pre-vacuum pump
Vac ready = OK
is dis-
played abnormally fast
Penning gauge has not been
identified correctly
Restart the microscope
If this does not solve the prob-
lem, contact your local ZEISS
service representative
Gun vacuum is worse
than 2 × 10⁻⁸ mbar
The pumping capacity of the
ion getter pump decreases in
the course of time, thus deteri-
orating the gun vacuum
Refer to
Tab. 6: Troubleshooting
7.1 Chamber
7.1.1 Initializing the Stage
If a stored stage position cannot be approached or if the stage does not move or does not move
accurately, the stage needs to be initialized.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4
Always close the chamber door before moving the specimen stage.
4
To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
Prerequisite
ü
The specimen chamber has been evacuated, refer to
Loading the Specimen Chamber [
ü
Requires the
Stage Initialise
privilege