Connections, establishing
Connecting the S100 floor stand
•
Ensure that the rated voltage indicated in the display window (3) (only
for halogen / xenon) corresponds with the rated voltage for the network
at the site of the installation.
•
Ensure that the unit has been switched off at the power switch before
plugging in or unplugging connectors at the power input socket (4).
•
Make sure that the power switch (1) is OFF.
•
Plug the power cord which is supplied with the supply unit into the power
inlet socket (4).
•
Connect the power cord via a socket to the power supply system.
CAUTION
Risk of electric shock caused by unapproved or defective accessories!
Unapproved or defective accessories can cause increased leakage current!
•
Do not touch any video outputs or other signal interfaces while you
are in contact with the patient.
•
When configuring your system, comply with the requirements of
IEC 60601-1, chapter 16; in some countries IEC 60601-1-1 is still valid.
•
Comply with the Medical Device Operator Ordinance or other national
laws.
WARNING
Danger! Electrical voltage!
•
To avoid the risk of electrical shock, this device may only be connected
to a power supply network which is provided with a protective ground
conductor.
NOTE
Tripping hazard!
Inappropriately laid cables present an increased risk of tripping.
•
Always route cables in such a manner that they do not obstruct you in
your work.
Summary of Contents for S100
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Page 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Page 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Page 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Page 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Page 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Page 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Page 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Page 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Page 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...