5 First Operating Steps | 5.7 Adjusting the Transmitted-Light Phase Contrast or the Transmitted-Light
Darkfield using the Turret Condenser
ZEISS
5.7 Adjusting the Transmitted-Light Phase Contrast or the Transmitted-Light
Darkfield using the Turret Condenser
Prerequisite
ü
The phase contrast method on the microscope requires a stand with Full-Köhler equipment..
ü
The turret condenser with phase contrast stop or darkfield stop inserted is required.
Procedure
1. Adjust the microscope as you would for transmitted-light brightfield, see
Transmitted-Light Brightfield on the Full-Köhler Microscope [
2. Turn the nosepiece to move the phase contrast
objective (
Ph 1
or
Ph 2
or
Ph 3
) into the light
path.
3. Open the luminous-field diaphragm on the
stand.
4. Turn the turret disk of the turret condenser to
the phase contrast position (
Ph 1
or
Ph 2
or
Ph 3
) that corresponds to the objective being
used.