5 First Operating Steps | 5.4 Adjusting the Transmitted-Light Brightfield on the Full-Köhler Microscope
6. Focus the image for the second eye by turning
the focusing ring of the second eyepiece.
9
Having done so, both microscopic images inclusive of the eyepiece pointer or micrometer are
focused.
Info
Afterwards, you should focus on the specimen using the focusing drive only.
5.4 Adjusting the Transmitted-Light Brightfield on the Full-Köhler Microscope
Parts and Tools
High contrast specimen slide with cover glass of 0.17 mm thickness
Info
The knurled knob for the vertical condenser adjustment is always on the side opposite to the
stage controls.
Prerequisite
ü
The microscope is adapted to the user, see
.
ü
A 10x objective for use with cover glass of 0.17 mm thickness is mounted.
Procedure
1. Rotate the 10x objective into the light path us-
ing the knurled ring of the nosepiece.
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en
69