5 First Operating Steps | 5.6 Adjusting the Transmitted-Light Phase Contrast or the Transmitted-Light
Darkfield using a Contrast Slider
Procedure
1. Adjust the microscope as you would for transmitted-light brightfield, see
Transmitted-Light Brightfield on the Full-Köhler Microscope [
2. Turn the nosepiece to move the phase contrast
objective (
Ph 2
) into the light path.
3. Open the luminous-field diaphragm on the
stand.
4. Remove the dummy slider from the Abbe con-
denser.
5. Remove the locking screw from the phase con-
trast slider (fitting the objective used,
Ph 2
).
Push the phase contrast slider from the left into
the Abbe condenser until it reaches the lock-in
position.
6. Screw the locking screw into the contrast slider.
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en