Failure of the magnetic brakes
If the magnetic brakes fail (magnetic brakes are locked), you can manu-
ally move the axes of the stand by overcoming the locking effect of the
brakes.
Touchscreen failure
Warning!
Do not under any circumstances touch the touchscreen. There is a possi-
bility that only the illumination of the display has failed. In this case, you
would inadvertently call menus or change settings when touching the
screen.
Note:
If the touchscreen no longer responds to your entries, you can still con-
tinue to operate the OPMI Pentero with the current settings.
To restore the touchscreen function, proceed as follows:
•
Switch the suspension system off and back on again after a short time
(approx. 2 minutes).
After power-up, the system automatically goes into the standard oper-
ating mode after it has completed a self-test (approx. 30 seconds).
•
Check all functions of the surgical microscope and suspension sys-
tem.
You can continue surgery in the standard operating mode.
Failure of the line voltage
In the event of power failure, the uninterruptible power supply (UPS) au-
tomatically starts to operate. It ensures for a short period of time that no
data will be lost.
Note:
The UPS does not power the light source and microscope functions.
Summary of Contents for OPMI Pentero
Page 7: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 7 3 4 5 6 5 6...
Page 25: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 25 BNC 1 Y C 2...
Page 29: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 29 1 2 3...
Page 39: ...G 30 1702 en OPMI Pentero Software Release 2 2 Issue 2 0 Printed on 17 03 2008 39...