What to do in the event of malfunctions
OPMI Lumera i
Version 8.0
Page 132
G-30-1720-en
Troubleshooting
For your safety
•
Many malfunctions are detected automatically. Therefore, pay attention
to the error messages displayed on the 5.7" control panel (see Page 135).
•
If there is a failure and you cannot resolve the issue using information in
the chapter “What to do in the event of malfunctions”, attach a sign “Out
of order” to the system and contact a ZEISS service representative.
Faults in the system
Fault
Possible reason
Corrective measure
Cross
reference
No function at all.
Power plug of stand not
inserted.
Plug in the power plug.
Power switch of stand not
switched on.
Press power switch. Green
indicator light in the power
switch must be lit.
Automatic circuit breaker in
the stand power switch is
being triggered.
Press power switch again.
Power failure.
Contact in-house electrician.
-
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...