OPMI Lumera i
Operation
Powering the device up and down
Prerequisites
The correct rated voltage has been set on the suspension system (see
Page 48).
The power cable has been connected (see Page 76).
The optical fiber is connected (see Page 72).
The foot control panel is connected (see Page 77).
Proceed as follows to power the system up and down:
•
Press the power switch (1) on the suspension system.
The system is switched on if the green lamp in the power switch is lit.
After power-up, the system performs a self-test.
During this time (approx. 120 seconds) a start screen with the Zeiss
logo is displayed on the 5.7" control panel.
Switching on the surgical microscope for the fist time displays the default user
profile “Default User”. In the next logins, the system displays the last user
selected.
The Default User profile can be neither edited nor changed.
Before starting to work with the system, you must perform a function test (see
next page). During this test, you can check the device configuration and user-
specific settings and reconfigure them, if necessary.
CAUTION
Increased leakage current!
There is a risk of user and patient sustaining an electric shock.
•
Never touch the contacts of plug connectors during contact with the
patient.
•
Do not connect any defective or unapproved accessories to the contacts
of connectors.
CAUTION
Risk of injury caused by defective components!
If the software control fails, you may not work with the device for safety
reasons.
•
Contact ZEISS Service.
1
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...