Sterilization
The Asepsis Sets available from ZEISS contain rubber caps and handgrips
which can be sterilized in autoclaves. For detailed information on sterilization,
please see the enclosed instructions “Preparation of resterilizable products”
for the respective Asepsis Set.
You can also use disposable sterile drapes to cover the system.
•
When draping the system, make sure there is enough slack in the drapes
to allow for movement of the microscope carrier and surgical microscope.
•
In order to ensure that the lamps are able to cool sufficiently and to avoid
any loss in lamp functionality, take care not to cover the ventilation slots.
Disinfection
It may be necessary to disinfect the surfaces.
The maximum concentrations for application are:
–
For alcohol (tested with 2 propanol): 60 %
–
For aldehyde (tested with glutaraldehyde): 2 %
–
For quaternary compounds (tested with DDAC): 0.2 %
NOTICE
Surface damage caused by use of wrong disinfectants!
Performing disinfection with the wrong disinfectants may result in damage
to the surfaces of the device.
•
Use an aldehyde and/or alcohol-based disinfectant. The addition of
quaternary compounds is acceptable.
•
In order to prevent surface tensions, you may only use the disinfecting
components specified below.
Summary of Contents for OPMI Lumera i on floor stand
Page 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Page 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Page 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Page 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Page 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Page 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Page 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Page 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...