Instructions for Use
3 Description of the device
OPMI LUMERA 700
3.5 Functional description
G-30-1673-en - 14.0 - 2018-07-12
83 / 236
3.5.6 RESIGHT functions
The device can be equipped with the RESIGHT 500 or RESIGHT 700
contact-free fundus viewing systems. The fundus viewing system
provides a detailed view of the retina without having to move the
microscope. Swiveling the fundus viewing system into the surgical
microscope's beam path inverts the microscope image, turning the
image upside down. The Invertertube E can be used to
automatically display the image correctly again. The following
RESIGHT 700 settings can be adjusted for automatic focus and
image inversion:
Function
Effect
Invertertube E
image inversion
Displays the image in an upright position (normal) or
inverts it.
RESIGHT 700
internal focus
speed
Controls the speed of the RESIGHT 700's internal
focus. The internal focus ensures that the surgical
microscope no longer needs to be vertically moved.
When working on the anterior segment of the eye,
the RESIGHT 700 can be moved out of the beam path
without blurring the surgical microscope image.
RESIGHT 700
internal focus
two-step speed
Switches the two-step speed on or off.
• If two-step speed is switched off: the device
focuses using the same speed set by the user.
• If two-step speed is switched on: tapping and
holding down the focus setting changes the
focus speed, starting from the smallest speed
value (factory setting) and then increasing in
speed.
3.5.7 DeepView
DeepView optimizes the depth of field or the light transmission of
the microscope image.
Function
Effect
DeepView is
switched on
Sharp depth perception at every magnification.
DeepView is
switched off
Optimal light transmission for a bright image with
high resolution.
3.5.8 Reset options
The reset options are basic settings that the device reverts back to
when the XY reset button is pressed on the OPMI or the device is in
the park position. The following functions can be adjusted:
Function
Effect
Reset XY
position
Moves the XY coupling at maximum speed to its
center position.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...