5 Daily startup
Instructions for Use
OPMI LUMERA 700
5.3 Positioning the device in the OR
122 / 236
G-30-1673-en - 14.0 - 2018-07-12
5.3.5 Orienting the integrated 22" monitor
1
3
2
Figure 71: Orienting the integrated 22" monitor
1
22" monitor
2
Suspension arm
3
Support arm
CAUTION!
Risk of crushing caused by moving parts!
Fingers may be crushed between the monitor and support arm.
u
Stay clear of the area between the suspension arm and the
monitor while the user terminal is in motion.
CAUTION!
Risk of crushing caused by lowering of the 22" monitor!
A loss of gas in the suspension arm may cause the integrated 22"
monitor to lower, causing injury to the patient.
u
Compensate for the loss of gas by readjusting the gas spring.
[
Action
1. Swivel the support arm and the suspension arm into the
desired horizontal position.
2. Tilt the suspension arm upward or downward until the desired
height is reached.
3. Hold on to the upper corner of the integrated 22" monitor and
set the desired viewing angle.