Installation Instructions

METROTOM

Summary of Contents for METROTOM

Page 1: ...Technical Services Installation Instructions METROTOM ...

Page 2: ......

Page 3: ...ipulate further conditions which must be fulfilled as a prerequisite for the notice of ap proval According to 18 of the RöV the METROTOM must be inspected by an authorized expert after 5 years of operation at the latest recurring inspections The responsible supervisory authority of the relevant state is usually the Trade Supervisory Authority the Occupational Health and Safety Administration or th...

Page 4: ...wall 2 3 Transport paths Check the floor and ceiling loading capacity of all transport paths leading to the installation site The doorways and trans port paths must be at least 10 cm wider and higher than the METROTOM assemblies to be transported including the transport vehicle The dimensions and weights of the METROTOM assemblies are specified in this text 2 4 Ambient conditions MetroTomography s...

Page 5: ...hat sufficient illumination is present The use of heat sources near the METROTOM must be avoided 2 5 General protective measures The machine must be protected against direct contact with dirt and grime e g casting sand metal filings oil coolants lubricants and soot Such deposits result in higher costs for cleaning and care and may cause malfunction of and or dam age to the CT scanner Particle conc...

Page 6: ...ike to have this service performed for you contact Carl Zeiss in Oberkochen In any case the measured result must be made available 2 7 Floor loading capacity We recommend that you consult a structural engineer for your statical calculations and any required floor reinforcements foundation etc The floor load exerted by the machine is specified in this brochure Page 13 10 1 0 10 20 30 40 50 100 1000...

Page 7: ...t connection is provided for the system 16 A This connection must be protected with an automatic cutout Alternatively you can also use other protective devices with an equivalent rating It is recommended to install a lightning current arrester e g Phoenix Flashtrab FLT 25 400 as a coarse protection against overvoltages We recommend making the connection according to the following connection diagra...

Page 8: ...able duct 3 6 Stabilization of the power supply If brief fluctuations occur in the root mean square value of the line voltage a line voltage analysis must be performed Contact us if you require this service We will be glad to advise you and recommend suitable measures A separate stub cable from the main distributing frame to the connecting point of METROTOM is sufficient in most cases 3 7 Machine ...

Page 9: ...loading and transport of packaged units may be performed only with hoisting rings or pallets the insulated pa kkaging must not be damaged in the process Please ensure that suitable transport vehicles operated by qualified personnel are provided for unloading the crates and transporting them to their storage site The type of transport vehicle used depends on local conditions The load carrying capac...

Page 10: ...Installation Instructions Page 10 3 14 Moving the METROTOM to another location To prevent damage to the machine our system specialists should be consulted before it is moved to another location ...

Page 11: ... L W H Height of secured unit max Sea air crate Truck pallet with sea air crate with pallet f road 1 Pallet for CT base assembly 3320 2270 460 2600 3 5 410 4910 2 Pallet for CT cover 3320 2400 430 2600 3 5 415 3915 3 Control cabinet on European pallet approx 350 kg 4 Computer and accessories in Palbox large Data are not available for the transport crates Note All values not entered are not current...

Page 12: ...stributed over the base of the machine jacks of 95 cm2 per jack Maximum difference in level of the 4 contact surfaces in relation to each other 4 mm Maximum inclination per contact surface 1 mm m c Path along which the rollers of the cover run d Free space required for maintenance e Free space required for initial installation of the METROTOM see also illustration above Note 5 Space requirement fo...

Page 13: ...ded 1350 mm o The shaded area must be accessible for servicing so that bulky and heavy components can be removed and installed if necessary Transport to the left or right also must be possible p Cable METROTOM control cabinet approx 5 m cable entry from below q Cable control cabinet computer station approx 13 m cable entry from below r Cable METROTOM computer station approx 20 m s Cable Network co...

Page 14: ...astening hole Pos No Radius R Angle α 1 1 55 mm 60o 4 1 90 mm 60o 1 2 180o 4 2 180o 1 3 300o 4 3 300o 2 1 60 mm 0o 5 1 107 5 mm 0o 2 2 120o 5 2 120o 2 3 240o 5 3 240o 3 1 72 5 mm 0o 6 1 125 mm 60o 3 2 120o 6 2 180o 3 3 240o 6 3 300o α 0 R 5 1 3 1 2 1 2 3 2 2 3 3 5 3 3 2 5 2 6 1 4 1 1 1 1 2 4 2 6 2 6 3 4 3 1 3 Ø 300 ...

Page 15: ......