Errors occurring prior to measuring operations
8-2
61211-8140502
METROTOM 1500 G3 1.4
Errors occurring prior to measuring operations
If the CT system cannot be put into operation, start by checking whether
the following errors can be ruled out:
Error
Cause
Action
Check
The CT system cannot
be put into operation
Wrong start-up se-
quence
–
Observe the specified se-
quence.
–
First, switch on the CT
system and then the com-
puter.
Power supply
The
main switch
is not
on.
Connect power supply:
–
Turn the
main switch
to
the «1» position.
Position «1» must
be visible.
Controller
The controller is off.
Switch the controller on:
–
Turn the
rotary switch
clockwise.
The indicator lamp
in the rotary
switch must be lit.
Drives
The drives are off.
Switch the drives on:
–
Press the two
buttons
of
the loading door simulta-
neously.
–
Afterwards press the
but-
ton
for the drives.
The indicator lamp
in the rotary
switch must be lit.
The
EMERGENCY STOP
button
is locked.
Unlock the button:
–
Slightly turn the
EMER-
GENCY STOP button
.
The button pops
out a few millime-
ters.
Operating mode
The operating mode has
not been enabled.
Set the operating mode:
–
Turn the
key switch
to the
«AUTO» position.
X-ray system
The X-ray system has
not been switched on.
Switch the X-ray system on:
–
Turn the
key switch
to
«ON».
The X-ray system is in
standby mode.
Door control
The door control has not
yet been initialized.
Perform the initialization
run.
–
See
➤
The indicator
lamps of the
«OPEN door» and
«CLOSE door» but-
tons do not light
up.
Summary of Contents for METROTOM 1500 G3
Page 8: ...6Table of contents 6 61211 8140502 METROTOM 1500 G3 1 4...
Page 18: ...Warranty 1 6 61211 8140502 METROTOM 1500 G3 1 4...
Page 72: ...Start up checklist 6 14 61211 8140502 METROTOM 1500 G3 1 4...
Page 104: ...Inspection measures 9 10 61211 8140502 METROTOM 1500 G3 1 4...
Page 110: ...2Glossary 2 61211 8140502 METROTOM 1500 G3 1 4...
Page 120: ......