Door control
Initializing travel of the loading door
Once the CT scanner has been switched on, you must carry out an ini-
tializing travel of the loading door.
Note:
The loading door can be moved only if the drives are on.
Initialization after cold start
Button and indi-
cator lamp
Action
Effect
Comment
1
Switch on the drives.
Once the drives have
been switched on, the
«CLOSE door» button
flashes quickly.
The green indicator
lamp is lit.
2
Press the «CLOSE
door» button and
keep it pressed until
the loading door is
completely closed.
The loading door is
closed.
3
Press the «OPEN
door» button and
keep it pressed until
the loading door is
completely open.
The loading door is
open. The reference
points for the loading
door are set.
The «CLOSE door»
button goes out.
Now, the loading door can be operated in normal mode. To move the
loading door, only press the corresponding button, it is not necessary to
hold down the button. The corresponding button flashes during the
travel movement.
NOTE
The travel movement of the loading door can be stopped using one of
the two buttons at any time. A new travel movement will be carried out
only after pressing one of the two buttons again.
Summary of Contents for METROTOM 1500 G3
Page 8: ...6Table of contents 6 61211 8140502 METROTOM 1500 G3 1 4...
Page 18: ...Warranty 1 6 61211 8140502 METROTOM 1500 G3 1 4...
Page 72: ...Start up checklist 6 14 61211 8140502 METROTOM 1500 G3 1 4...
Page 104: ...Inspection measures 9 10 61211 8140502 METROTOM 1500 G3 1 4...
Page 110: ...2Glossary 2 61211 8140502 METROTOM 1500 G3 1 4...
Page 120: ......