Mechanical stage S mot. CAN
SET-UP
Measuring stage S mot. CAN
Mounting the mechanical / …
Carl Zeiss
08/2012
435465-9000-701
17
3.2.3
Mounting the mechanical stage /
measuring stage S mot. CAN on the
transillumination base 300
•
If necessary, remove the glass plate (Fig. 7/
1
)
from the transillumination base 300 (Fig. 7/
2
) by
pressing on the rear edge to tilt it.
•
Insert stage adapter 84+120 (Fig. 7/
3
) into the
transillumination base 300 and screw it hand-
tight with the 4 captive screws.
•
Set the mechanical / measuring stage (Fig. 7/
5
)
on the transillumination base 300 (stage drive
pointing to the left), align to the rear edge and
secure with the three captive screws (Fig. 7/
4
) of
the stage.
•
Insert the glass plate (Fig. 7/
6
) into the stage.
The procedure for inserting the plate
into and removing from the stage is
described on page 15.
The red dots on the table (Fig. 7/
5
) and
metal plate (Fig. 7/
6
) are to ensure
reproducibility of the alignment of the
metal plate and stage if the former
needs to be detached.
Fig. 7
Mounting the
mechanical /
measuring stage mot.
on
transillumination base 300