CHAPTER 3 - ADDITIONAL SOFTWARE TOOLS
ZEISS
SystemMaintenance Tool
LSM 880
4
000000-2071-464
10/2014 V_01
2.2
Starting the Maintenance Tool
Before starting the Maintenance Tool, the system hardware must be fully activated.
To do this, open ZEN and check the following settings: Correct entry of all existing lenses; the calibration
objective needs to be entered as
APO Calibration LSM
.
•
Correct entry of all existing lenses; the calibration objective needs to be entered as
APO Calibration
LSM
.
•
For LSM systems with a 405 nm laser, an approximate collimator position should be saved for the
calibration objective.
•
The focus plane needs to be approximately correct
•
All light sources (Hal, LED’s, lasers) need to be activated
•
If all these settings are correct, close ZEN and then start the
SystemMaintenance Tool
by clicking the desktop icon (Fig. 5).
During the system initialization phase, the SystemMaintenance Tool will
display the boot status (Fig. 6).
•
After start, choose
Log in… Customer
(type-in any name.
Fig. 5
Desktop icon of the
Maintenance Tool
Fig. 6
Boot status of the Maintenance Tool
Fig. 7
Log in window
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......