LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Display and Analysis of Images
Carl Zeiss
03/06
B 45-0021 e
4-341
Linear Unmixing
button
Performs the unmixing with the selected spectra.
Apply to hardware
button
Sets the parameters for the META detector used in binning mode as
specified in the settings ChS1, ChS2, ChS3 ... ChS8. See
Single Track
and
Multi Track
in
Config Control
for the effect of the action. The
Lambda Stack
setting is not affected by this action.
Save to spectra DB
button
The values of the intensity Lambda diagram are saved in a file
(extension: *.rls) in the aim/macro/roi directory. Make sure to have
displayed only data of single dyes in the intensity Lambda diagram when
executing this function.
4.13.28.4 Functionality for Emission Finger-
printing
Extract Channels panel
−
Sum signals from selected detector elements
−
Mean signals from selected detector
elements
−
Autoscaling of output channels (individually)
Linear Unmixing panel
−
Autoscaling of output channels
−
Generate an additional "Channel with
Residuals"-> pixel-by-pixel display of the
difference between fit and original data (for
the channel of the Lambda Stack that shows
the greatest deviation)
−
Advanced Linear Unmixing ignored negative
results and repeats the calculation.
−
Ask for ROI characterizing background
spectra before Linear Unmixing
−
Ignore Overexposed and Underexposed
Pixels. Those wont be considered for the
calculation.
Context menu
−
Spectra Display
with option
Normalize
(right mouse click into spectra graph)
Fig. 4-316
Extract and Unmix Properties window
Fig. 4-317
Context menu with Normalize function