OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Display and Analysis of Images
LSM 510 META
DuoScan
4-304
B 45-0021 e
03/06
(2)
Changing the topography geometry
The three buttons in the
Geometry
button bar allow the surface geometry to be changed.
Inverse
button
Inverse surface. Depths change to heights, and vice versa.
Fit
button
The following fit modes can be set:
1) No Fit
2) Plane
The topography is tilted in such a way that the mean deviation value
plane is calculated.
3) Manu Tilt
4) 3Point Tilt
5) Sphere fit correction
A spherical form is eliminated, determination of micro roughness on
spherical surfaces can be performed.
6) Cylinder fit correction
A cylinder form is eliminated, determination of micro roughness on
cylindrical surfaces can be performed.
You can display the exact values of the
Cylinder / Sphere
fit by
opening a context menu in the
Image Display
box with a click of
the right mouse button and selecting the
Show processing
parameter
function.
Manu Tilt
button
Clicking on the
Manu Tilt
button opens the
Manual Tilt
window
(Fig. 4-291) for manual tilt correction.
In the manual tilt window three types of arrow buttons are present:
−
One arrow changes pitch and yaw in 0.001° steps
−
Two arrow changes pitch and yaw in 0.01° steps
−
Three arrow changes pitch and yaw in 0.1° steps
•
Click on the
Inverse
button for the inverse display of the topography. Clicking again will reset the
normal display.
•
Correct the tilt via the
Fit
or
Manu Tilt
functions.
Fig. 4-291