NOTES ON DEVICE SAFETY
LSM 510
DuoScan
Carl Zeiss
User Interface
LSM 510 META
DuoScan
1-26
B 45-0021 e
03/06
2. The plug for the main connection of the
microscope to the electronics is situated on
the rear side of the electronics rack. It is either
of the four CAN connections shown in Fig.
1-25. Only ONE microscope can be connected
to the electronics at a time. The connection
has to be plugged in and unplugged at the
electronics rack. The corresponding
connections on the microscopes are shown in
Fig. 1-23/
1
1
. Remove the
connection of the microscope no longer in use
at the electronics rack and plug in the
connection of the microscope which should be
used instead.
3. After the connections of the non used
microscope are disconnected and the ones of
the used microscope are connected, the
system can be switched on again. Before
initializing the system with the LSM Software
make sure to use the right database according
to the microscope in use. It can be chosen via
the icon Stand Select.
Fig. 1-26
CAN connections on the
Axiovert 200 M (1). One of
the plugs is occupied with the
connection to the electronics
Fig. 1-27
The Axioskop 2 FS
MOT
is connected
to the electronics via the Interface
Control. The CAN connection is
fixed (1) to the control box