OPERATION IN EXPERT MODE
LSM 510 META
Maintain Menu
Carl Zeiss
03/06
B 45-0021 e
4-229
•
The calibration standard has to be positioned as indicated in Fig. 4-226.
•
Before starting the calibration procedure, change to the bidirectional scan mode in the
Scan Control
window.
•
Scan Corr. X
and
Scan Corr. Y
in the
Mode
sheet of the
Scan Control
window has to be set to
zero.
•
Click on the
Scanner
button in the
Maintain
subordinate toolbar of the
Main
menu.
−
This opens the
Scanner Calibration
window.
•
Click on the
Speed 11-13
(optical) button.
•
Activate the
Display Graphics
check box enables the graphical display of the calibration process.
•
Click on the
More
(Less) button to display the Speed Selection menu.
Fig. 4-226
Correct orientation of calibration sample for X scanner calibration