3.3.4
Insertion of Accessories for Incubation
•
A
temperature sensor
1
)
can be used in the sample chamber body.
−
Hold the temperature sensor so that it shows you the letter “J”. Slide the temperature sensor
sleeve (Fig. 23/
3
) onto the top horizontal part of the J-shaped temperature sensor.
−
Dip the temperature sensor with sleeve into the sample chamber body and position the
temperature sensor sleeve securely into the sample chamber body groove (Fig. 23/
2
).
−
Place a
cover
on top of the sample chamber body (Fig. 24/
3
). This will minimize evaporation of
medium or contamination of the sample chamber. If CO
2
module Lightsheet Z.1 is used
(configuration dependent, chapter 2.7.2) the cover is recommended to maintain the CO
2
gas
concentrations.
−
Keep the opening as small as possible. Use the 3 mm opening cover for capillaries (Fig. 24/
1
) and
7 mm opening cover for syringes (Fig. 24/
2
). The cover only rests on top of the sample chamber
body and is free to move.
1
Temperature sensor
2
Sample chamber body groove
3
Temperature sensor sleeve
Fig. 23
Insertion of temperature sensor