3.3
Assembly of the Sample Chamber
−
Wear powder-free gloves to avoid fingerprints on the cover slips as well as to protect your hands
from sharp edges.
3.3.1
Assembly of the Sample Chamber Windows
−
Assemble the three sample chamber body windows (for the
illumination optics on the sides
and
the front chamber window for the
overview camera with LED white illumination
) using the
following steps:
1.
Take the sample chamber body (Fig. 18/
1
) and lay it onto the table so that one of the three smaller
windows are facing you.
2.
Using a forceps or gloved hands, remove one clean circular 18 mm cover slip; place cover slip into
the empty sample chamber body window so it fits into the smallest groove (Fig. 18/
2
) – you will
need a fine forceps.
3.
Now place the O-ring (18 mm O-ring, black) into the corresponding groove (Fig. 18/
3
). Make sure
not to disturb the positioning of the cover slip.
4.
Take the silver illumination adapter ring (Fig. 18/
4
) and lay it into the window making sure the flat
side with the four pinholes is facing you.
5.
Using the sample chamber window tool (Fig. 18/
5
), position the circular pins into the pinholes and
turn clockwise until finger-tight. Adapter should be flush with the sample chamber body.
1
Sample chamber body
2
Cover slip
3
O-ring (18 mm O-ring, black)
4
Illumination adapter ring
5
Sample chamber window tool
Fig. 18
Assembly of the three sample chamber windows for illumination optics and overview
camera