Lightsheet Z.1
Left Tool Area and Hardware Control Tools
Carl Zeiss
02/2013
000000-1790-528
75
3.2.11.4
End
Expand the
End
field by clicking the triangle
In the
Mode
drop down menu (Fig. 115),
you can select the way the acquisition is
triggered. You have the choice between
Manual and Time.
If
Manual
is selected the acquisition of a
time series ends using the
Stop button
or
after the defined number of cycles.
If
Time
is selected the
hh
(hour),
min
(minute) and
sec
(second) input boxes
become available (Fig. 116), in which the
hours, minutes and seconds can be edited.
The system interrupts the acquisition after
the defined computer time has been
reached.
A Time Series is finished when the end Time is reached.
If the entered number of Cycles has been processed, the Time Series is finished before the end
Time. If the number of cycles has not yet been processed until the end Time, the Time Series is also
finished. The end time for the Time Series can be changed online.
If a time series is interrupted before its programmed end, the programmed number of images will
be taken over in the database. However, only those images are stored which were created before
interruption of the time series. This is due to the fact that the original image parameters are to be
taken over via the
Reuse
function.
Fig. 114
End field
Fig. 115
End Mode drop down menu
Fig. 116
End Mode drop down menu – Time selected