SETUP REQUIREMENTS
Lattice Lightsheet 7
Environmental Requirements
ZEISS
04/2021 V_01
000000-2324-960
63
2.6
Environmental Requirements
For operation the system must be placed in a closed room.
1. Operation, specified performance without
incubation
T = 22 °C ±0.5 °C without interruption (24 h a day
independently whether system is operated or
switched off)
It has to be ensured that the air-flow of the air-
conditioning is not directed at the system.
2. Operation, reduced performance without
incubation
T = 15 °C to 30 °C, any conditions different from
item 1. and 5.
3. Storage, less than 16 h
T = 4 °C to 30 °C
4. Storage, less than 6 h
T = -20 °C to 55 °C
5. Temperature gradient
±0.5 °C/h
6. Warm-up time
1 h;
≥
3 h for high-precision
and/or long-term measurements
7. Relative humidity
< 65 % at 30 °C
8. Operation altitude
max. 2000 m
9. Heat emission
0.8 kW
2.7
Vibrations
According to ISO 10811 operation of the Lattice Lightsheet 7 system is recommended for vibration class C
(VC-C, 12.5 µm/s for 8-80 Hz (RMS)).
2.8
Laser for Lattice Lightsheet 7
Single mode polarization preserving fiber
Laser beam attenuation for all lasers by modulation
•
Laser line 488 nm, 10 mW (at output), laser class 3B
•
Laser line 561 nm, 10 mW (at output), laser class 3B
•
Laser line 642 nm, 5 mW (at output), laser class 3B