Humphrey Matrix Visual Field Instrument
Rev. D 08/21/06 PN 701692
Recommended separation distances between portable and mobile RF communications
equipment and the Humphrey Matrix
The Humphrey Matrix is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Humphrey Matrix can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Humphrey Matrix as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d=1.17
√
P
V1
d=1.17
√
P
E1
d=2.33
√
P
E1
0,01
0.117
0.117
0.233
0,1
0.370
0.370
0.737
1
1.170
1.170
2.330
10
3.700
3.700
7.368
100
11.700
11.700
23.300
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Summary of Contents for Humphrey matrix
Page 25: ...Humphrey Matrix Visual Field Instrument 21 Rev D 08 21 06 PN 701692 Figure C...
Page 63: ...Humphrey Matrix Visual Field Instrument 59 Rev D 08 21 06 PN 701692...
Page 64: ...60 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 65: ...Humphrey Matrix Visual Field Instrument 61 Rev D 08 21 06 PN 701692...
Page 66: ...62 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 67: ...Humphrey Matrix Visual Field Instrument 63 Rev D 08 21 06 PN 701692...
Page 68: ...64 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 69: ...Humphrey Matrix Visual Field Instrument 65 Rev D 08 21 06 PN 701692...
Page 70: ...66 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Page 71: ...Humphrey Matrix Visual Field Instrument 67 Rev D 08 21 06 PN 701692...
Page 72: ...68 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...